8 WA1K2K-OM-E RevB
Fuse Replacement
(For Only WaveAce 2000 Models Only)
Set the instrument Standby (power) switch to Standby mode, flip the
mains power switch to the OFF (0) position, and disconnect the power
cord before inspecting or replacing the fuse. Open the black fuse holder
(located at the rear of the instrument below the main power switch)
using a small, flat-bladed screwdriver. Remove the old fuse, replace it
with a new 5 X 20 mm T rated 1.25 A/250 V fuse, and reinstall the fuse
holder.
WARNING
For continued fire protection at all line voltages, replace fuse with the
specified type and rating only. Always disconnect the power cord before
replacing the fuse.
Calibration
The recommended calibration interval is one year. Calibration should be
performed by qualified personnel only.
Cleaning
Clean only the exterior of the instrument, using a damp, soft cloth. Do
not use chemicals or abrasive elements. Under no circumstances allow
moisture to penetrate the instrument.
Avoid electrical shock hazard by unplugging the power cord from the AC
outlet before cleaning.
WARNING - Electrical Shock Hazard
No operator serviceable parts inside.
Do not remove covers.
Refer servicing to qualified personnel.