Voltage (horizontal cursors) or Time (vertical cursors). Press the
Source option button and select a source of CH1, CH2 (also, CH3
and 4 on 4 channel models). Manual Mode also allows you to
choose MATH, REFA, or REFB as a Cursor source. Press the CurA or
CurB option button and turn the Adjust front panel knob to adjust
the cursors.
OR
Track
Set the mode to Track by pressing the Mode option button until
Track is selected. Press the Cursor A or Cursor B option button and
select a source of CH1, CH2 (also, CH3 and 4 on 4 channel models)
or NONE. Press the CurA or CurB option button and turn the
Adjust front panel knob to adjust the cursors. If Track is selected,
both cursors move in unison and appear slightly brighter on the
grid display.
OR
Auto
Set the mode to Auto by pressing the Mode option button until
Auto is selected. When you select measurement parameters using
the MEASURE front panel button, the cursors are automatically
displayed.
Note: When using Cursors, their measurement values are always shown in the
upper-left corner of the grid display. The cursor selected for placement is
indicated by a highlighted icon on the menu and the cursor itself has a higher
brightness on the grid display.
Parameter Measurements
Waveform analysis typically begins with the measurement of
parameters. Parameter measurement tools determine a wide range of
waveform properties. Use them to automatically calculate many of your
waveform attributes, like rise time, rms voltage, and peak-to-peak
voltage, for example.
Access parameter measurements by pressing the MEASURE front panel
button. The main Measure menu is shown. Notice how each row on the