User manual

Table Of Contents
Operator's Manual
70 WA1K2K-OM-E RevB
PLEASE NOTE THE FOLLOWING:
The Recall option is excluded from Save/Recall menus when
Picture or CSV is selected as the Type.
The Load option is disabled (grayed-out) on menus inside the
SAVE ALL screen when Pictures or CSV file types are selected as
Type.
Recalling Factory Settings
The oscilloscope is set up for normal operation when it is shipped from
the factory.
There are two ways to reload the factory settings:
1. Press the DEFAULT SETUP front panel button.
OR
2. Press the Save/Recall front panel button. The SAVE/REC menu
opens. Press the Type option button and select Factory.
3. Now, press the Load option button.
Applying the default setup does not reset the following:
Language option
Saved reference waveform files
Saved setup files
Display contrast
Calibration data