User manual

Table Of Contents
Operator's Manual
92 WA1K2K-OM-E RevB
Press the SINGLE front panel button to begin loading the .ads
firmware update file from your USB memory device Using the
SAVE ALL Screen (on page 74). Otherwise, press the RUN/STOP
front panel button to exit the Update Firmware screen.
Locate the .ads firmware update file, select it, and then press the
Load option button. The following status screen is shown.
The oscilloscope then beeps aloud twice and shows a screen
stating - The firmware has been updated completely, please boot
up the DSO again. At this point, turn your oscilloscope power Off,
wait a short while, and then power it On again.
When the oscilloscope has rebooted, your update is complete.