User Manual

Table Of Contents
Page 125 of 129 54A0028-6 042019 © LeddarTech Inc.
Amplitude (count): 47 CC hex = 18380 / 64 = 287.19 counts
Flag: 01 hex = 1 (Valid)
Segment #: 01 hex = 1 (Segment #2)
(8) 39 00 64 33 01 00:
Distance (cm): 00 39 hex = 57cm = 0.57m
Amplitude (count): 33 64 hex = 13156/ 64 = 205.56 counts
Flag: 01 hex = 1 (Valid)
Segment #: 00 hex = 0 (Segment #1)
Get detection messages (trailing fields):
(1) F0 4C 1A 00 64 00 00:
TimeStamp (ms): 00 1A 4C F0 hex = 1 723 632 ms = 1723 s
Light source POWER (%): 64 hex = 100 = 100%
Bit field acq. (reserved): 00 00 hex = 0
CRC (16bits Modbus) = EE D2