User Manual
Table Of Contents
- 1. Important Information
- 1.1 Symbols and their meanings
- 1.2 Qualification of personnel
- 1.3 Intended use of instrument
- 1.4 Instrument type
- 2. Safety
- 2.1 Safety notes
- 2.2 Warnings
- 2.3 Safety features on the instrument
- 3. Instrument Components and Specifications
- 3.1 Overview – instrument components
- 3.2 Specific instrument options
- 3.3 Standard delivery - packing list
- 3.4 Technical Data
- 3.5 Compatible reagents
- 3.6 Recommended reagent handling
- 3.6.1 Cycle for changing reagents
- 4. Instrument Setup
- 4.1 Installation site requirements
- 4.1.1 Moving the instrument
- 4.2 Electrical connection
- 4.2.1 Connecting the power supply
- 4.2.2 Connecting an uninterruptible power supply (UPS)
- 4.3 Installing accessories
- 4.4 Making the data connections
- 4.5 Anti-reflection clip – Function
- 4.6 Alarm functions
- 4.7 Switching the instrument on
- 4.8 Touchscreen functions
- 4.9 Checklist for initial operation
- 4.10 Switching off the instrument
- 5. Operation
- 5.1 Setting up the instrument parameters
- 5.1.1 System setup
- 5.1.2 Access levels
- 5.1.3 INSTALLATION menu
- 5.1.4 Editing the reagent list
- 5.1.5 Viewing the program list
- 5.1.6 Adding and/or modifying programs
- 5.1.7 Favorites
- 5.1.8 Configuring the stations
- 5.1.9 Reagent groups
- 5.2 Reagent handling
- 5.2.1 Draining/filling reagents (other than paraffin)
- 5.2.2 Replacing the paraffin
- 5.3 Running programs
- 5.3.1 Editing a program that has been called up.
- 5.3.2 Starting a program
- 5.3.3 Opening the retort
- 5.3.4 Retort emergency release
- 5.4 Reagent status
- 5.5 System monitor
- 5.6 The online help
- 6. Troubleshooting
- 6.1 General
- 6.2 Power failure
- 6.3 Troubleshooting
- 6.4 Typical fill or drain problems
- 7. Cleaning and maintenance
- 7.1 Clean programs
- 7.1.1 Retort cleaning programs
- 7.1.2 Paraffin cleaning
- 7.1.3 Smart clean
- 7.2 General cleaning steps
- 7.2.1 Daily cleaning and maintenance
- 7.2.2 Periodic cleaning and maintenance
- 7.3 Checklist for preventive maintenance
- 8. Warranty and Service
- Appendix 1
- Notes on Specimen Preparation

48
Leica ASP300 S
5.1.5 Viewing the program list
Youcancreate:
• Upto15infiltrationprograms
• 3retortcleaningprograms
• 1paraffincleaningprogram
At supervisor access level:
• Theprogramnamesfortheinfil-
tration programs can be edited.
• Newinfiltrationprogramscanbe
added and existing ones can be
deleted.
Fig. 43
→
This list (Fig. 43) shows all programs defined in the ASP300 S.
5. Operation
Important!
New infiltration programs are created by copying an existing pro-
gram. Therefore, the list must contain at least one program at all
times.
Retort or paraffin cleaning programs are preset. They cannot be
renamed, added or deleted.
Program duration cannot be specified. It is determined by the total duration
of all program steps, plus the estimated fill and drain times. To alter the
duration of a program, the duration of one or more individual program steps
has to be modified.