5. Operation
Remote draining
• Onthescreen,selectthereagentcontainer
(No. 3) and the receptacle (External).
Then press STATION DRAIN.
• Checkthatnoreagenthasleaked.
Remote filling
On thescreen,selectanempty reagent
container (No. 3) and the external receptacle
(External).
Then press the STATION REFILL button.
• Theremotefillingshouldbecompletedinless
than 170 seconds.
• Checkthatnoreagenthasleaked.
Select station Press button
Filling/draining reagents
From the FAVORITES screen, press MENU. In
the MENU FUNCTIONS window, tap the SMART
SCREEN button.
The SMART SCREEN (Fig. 53) pops up. The smart
screen is the initial screen for manual operation
of the instrument.
Fig. 53
Fig. 54
Status bar
displayed in supervisor mode.
Status bar
The following values are displayed.
• Pressureandtemperatureoftheretort
• Waxbathtemperature
• Valueofthelevelsensoroftheretort
• Retortlid–closedoropen
The following steps should only be
carried out by trained laboratory
personnel, experienced in handling