User Manual
Table Of Contents
- 1. Important Information
- 1.1 Symbols and their meanings
- 1.2 Qualification of personnel
- 1.3 Intended use of instrument
- 1.4 Instrument type
- 2. Safety
- 2.1 Safety notes
- 2.2 Warnings
- 2.3 Safety features on the instrument
- 3. Instrument Components and Specifications
- 3.1 Overview – instrument components
- 3.2 Specific instrument options
- 3.3 Standard delivery - packing list
- 3.4 Technical Data
- 3.5 Compatible reagents
- 3.6 Recommended reagent handling
- 3.6.1 Cycle for changing reagents
- 4. Instrument Setup
- 4.1 Installation site requirements
- 4.1.1 Moving the instrument
- 4.2 Electrical connection
- 4.2.1 Connecting the power supply
- 4.2.2 Connecting an uninterruptible power supply (UPS)
- 4.3 Installing accessories
- 4.4 Making the data connections
- 4.5 Anti-reflection clip – Function
- 4.6 Alarm functions
- 4.7 Switching the instrument on
- 4.8 Touchscreen functions
- 4.9 Checklist for initial operation
- 4.10 Switching off the instrument
- 5. Operation
- 5.1 Setting up the instrument parameters
- 5.1.1 System setup
- 5.1.2 Access levels
- 5.1.3 INSTALLATION menu
- 5.1.4 Editing the reagent list
- 5.1.5 Viewing the program list
- 5.1.6 Adding and/or modifying programs
- 5.1.7 Favorites
- 5.1.8 Configuring the stations
- 5.1.9 Reagent groups
- 5.2 Reagent handling
- 5.2.1 Draining/filling reagents (other than paraffin)
- 5.2.2 Replacing the paraffin
- 5.3 Running programs
- 5.3.1 Editing a program that has been called up.
- 5.3.2 Starting a program
- 5.3.3 Opening the retort
- 5.3.4 Retort emergency release
- 5.4 Reagent status
- 5.5 System monitor
- 5.6 The online help
- 6. Troubleshooting
- 6.1 General
- 6.2 Power failure
- 6.3 Troubleshooting
- 6.4 Typical fill or drain problems
- 7. Cleaning and maintenance
- 7.1 Clean programs
- 7.1.1 Retort cleaning programs
- 7.1.2 Paraffin cleaning
- 7.1.3 Smart clean
- 7.2 General cleaning steps
- 7.2.1 Daily cleaning and maintenance
- 7.2.2 Periodic cleaning and maintenance
- 7.3 Checklist for preventive maintenance
- 8. Warranty and Service
- Appendix 1
- Notes on Specimen Preparation

58
Leica ASP300 S
5.2.2 Replacing the paraffin
Draining the paraffin
• Connect theparaffindrainhose (36) to the
spout (17) of the paraffin drain at the front of
the instrument (Fig. 55) and into a catch tank.
Fig. 55
• Selecttheparaffinreservoirtobeemptiedand
the receptacle (External, Fig. 56).
Then, press STATION DRAIN....
• Thisisfollowedbymultiplesecurityprompts
that you have to confirm individually in order
to continue.
17
36
From FAVORITES, call up the SMART SCREEN
screen. (see previous chapter)
→
+
→
Fig. 56
5. Operation
When connecting the hose, be sure to
press it onto the O-rings of the drain
opening as far as it will go.
The paraffin drain hose must be
securely hooked into the external
receptacle and remain there through
the entire drainage process.
When drainage is finished, the hose is
cleaned automatically using air.
Do not remove the hose from the exter-
nal receptacle until this cleaning step
is complete.