User Manual
Table Of Contents
- 1. Important Information
- 1.1 Symbols and their meanings
- 1.2 Qualification of personnel
- 1.3 Intended use of instrument
- 1.4 Instrument type
- 2. Safety
- 2.1 Safety notes
- 2.2 Warnings
- 2.3 Safety features on the instrument
- 3. Instrument Components and Specifications
- 3.1 Overview – instrument components
- 3.2 Specific instrument options
- 3.3 Standard delivery - packing list
- 3.4 Technical Data
- 3.5 Compatible reagents
- 3.6 Recommended reagent handling
- 3.6.1 Cycle for changing reagents
- 4. Instrument Setup
- 4.1 Installation site requirements
- 4.1.1 Moving the instrument
- 4.2 Electrical connection
- 4.2.1 Connecting the power supply
- 4.2.2 Connecting an uninterruptible power supply (UPS)
- 4.3 Installing accessories
- 4.4 Making the data connections
- 4.5 Anti-reflection clip – Function
- 4.6 Alarm functions
- 4.7 Switching the instrument on
- 4.8 Touchscreen functions
- 4.9 Checklist for initial operation
- 4.10 Switching off the instrument
- 5. Operation
- 5.1 Setting up the instrument parameters
- 5.1.1 System setup
- 5.1.2 Access levels
- 5.1.3 INSTALLATION menu
- 5.1.4 Editing the reagent list
- 5.1.5 Viewing the program list
- 5.1.6 Adding and/or modifying programs
- 5.1.7 Favorites
- 5.1.8 Configuring the stations
- 5.1.9 Reagent groups
- 5.2 Reagent handling
- 5.2.1 Draining/filling reagents (other than paraffin)
- 5.2.2 Replacing the paraffin
- 5.3 Running programs
- 5.3.1 Editing a program that has been called up.
- 5.3.2 Starting a program
- 5.3.3 Opening the retort
- 5.3.4 Retort emergency release
- 5.4 Reagent status
- 5.5 System monitor
- 5.6 The online help
- 6. Troubleshooting
- 6.1 General
- 6.2 Power failure
- 6.3 Troubleshooting
- 6.4 Typical fill or drain problems
- 7. Cleaning and maintenance
- 7.1 Clean programs
- 7.1.1 Retort cleaning programs
- 7.1.2 Paraffin cleaning
- 7.1.3 Smart clean
- 7.2 General cleaning steps
- 7.2.1 Daily cleaning and maintenance
- 7.2.2 Periodic cleaning and maintenance
- 7.3 Checklist for preventive maintenance
- 8. Warranty and Service
- Appendix 1
- Notes on Specimen Preparation

64
Leica ASP300 S
5.3.2 Starting a program
Fig. 67
↓
→
→
Fig. 68
5. Operation
• PressingSTART starts the program. Confirm the prompt that follows with
OK. The number of blocks to be edited (NUMBER OF BLOCKS) must be
entered if the RMS has been enabled.
• TheFINISH TIME button displays the time when the currently running
program will end. If events occur that cause the end time to be delayed,
the end time display here will be updated accordingly.
• Tomodifyarunningprogramoropentheretorttoreloadspecimens(see
Chap. 5.3.3 "Opening the retort"), the program has to be stopped – press
the PAUSE button. All buttons that have changeable functions are acti-
vated and can be edited.
To change or delete program steps, press EDIT and make the changes
in the input window. CONTINUE restarts the program.
The end time of a running program can be changed only if the
delay step has not yet been completed.
• Youcanswitchfromthegraphicaltothetabledisplaybytouchingthe
TABLE, and back again with GRAPHICAL.