Instruction Manual
13
Leica CM1850 UV – Cryostat
Please refer to section 4.1 ‘Site require-
ments‘!
3. Technical data
All CM 1850 UV instrument types require the following secondary fuses:
F1: T0.25 A Fa. Schurter, Type FST; 6.3x32 mm or T0.25 A Fa. Littlefuse, Type 313; 6.3x32 mm
F2: T0.6 A Fa. Schurter, Type FST; 6.3x32 mm or T0.6 A Fa. Littlefuse, Type 313; 6.3x32 mm
F3: T1.6 A Fa. Schurter, Type FST; 6.3x32 mm or T1.6 A Fa. Littlefuse, Type 313; 6.3x32 mm
F4: T6.25 A Fa. Schurter, Type FST; 6.3x32 mm or T6.25 A Fa. Littlefuse, Type 313; 6.3x32 mm
F5: T4 A Fa. Schurter, Type FST; 6.3x32 mm or T4 Fa. Littlefuse, Type 313; 6.3x32 mm
* Refrigerant and compressor oil must only be replaced by qualified, authorized service per-
sonnel!
Microtome
Rotary microtome
Section thickness setting 1 - 60 μm
Total specimen feed 25 mm
Vertical stroke 59 mm
Maximum specimen size 55 x 55 mm
Specimen orientation 8° (x-, y-, z-axis)
Electric coarse feed
slow
200 μm/s
rapid 700 μm/s
Cryocabinet
Dimensions
Width (w/o handwheel) 600 mm
Width (with handwheel) 730 mm
Depth 730 mm
Height 1140 mm
Weight (incl. microtome,
without specimen cooling) ca.135 kg
Lamp
50 Hz version: Osram Dulux S 11 W/21
Color: LUMILUX light white
60 Hz version: Osram Dulux S 13 W/21
Color: LUMILUX light white
Storage conditions
Temperature: 5 - 55 °C
Relative humidity: < 85 %, non-condens-
ing