Leica CM1850 UV – Cryostat
9. Cleaning, disinfection, maintenance
• Loosen the screws (3) with an Allen key (4 mm).
Do not loosen the screws (7) for removing the
microtome.
• Disconnect the coarse feed motor plug (1) by
pulling the metal head.
• Disconnect the temperature sensor (2) from
the microtome.
1
2
3
4
5
7
8
8
• Slightly lift the microtome and pull it to the left
to disengage the plastic coupling (5) connect-
ing the axes.
• Take the microtome out of the cryochamber.
Fig. 25
Fig. 26