• Turn the instrument off with the automatic
mains fuse.
• Pull the mains plug.
• Slightly lift the sliding window (1) holding it by
the grip (2) and pull it out to the front.
• For the technical specification of the lamp,
please refer to Chapter 4. ‘Technical data‘.
Removal of the lamp
The lamp (3) is mounted behind a glare shield (4)
and therefore not visible.
• Touch the lamp for better orientation.
• Lightly tilt the fluorescent tube down to the left
and pull it out of the clip (5).
• Hold the lamp with both hands and pull it to
the left out of the holder (6).
Installation of the new lamp
• Hold the lamp in the correct mounting posi-
tion as shown (Fig. 35) and push it to the right
until it engages in the holder.
• Lightly push the fluorescent tube upward to
engage in the clip.
• Replace the sliding window.
• Reconnect the instrument to the mains power
and turn it on.
1
2
3
45
6
Fig. 34
Fig. 35
Fig. 36
9.3.7 Replacement of the lamp