9.3.3 Replacing the UVC lamp
Switch the instrument off and disconnect the power plug before replacing the fluorescent
lamp! Wear suitable safety gloves and safety glasses.
If the lamp is broken, it must be replaced by the technical service, as the replacement involves
a high risk of injury.
A UVC lamp has an estimated service life of approx. 9,000 hours.
Each on/off switching cycle reduces the lamp life by approx. one hour plus
burning time (30 minutes or 180 minutes respectively).
When both disinfection indicator lights (short and long-term dis-
infection) are flashing alternately in control panel field 1, the UVC
lamp must be replaced.
9.3.2 Replacing the fuses
9. Cleaning, Disinfection, Maintenance
On the back of the instrument there is a fuse box with 5 fuses:
• Unscrewthefusecapwithascrewdriver.
• Removebothfusecapandfuse.
• Insertthe definedfuseintothe sleeve, thenusethe
screwdriver to screw the sleeve back in as far as it will
go.
Fuse Protection Type
F1 Display T 0.25 A
F2 Coarse feed T 1.6 A
F3 Processor board supply T 1.0 A
F4 Peltier element T 6.25 A
F5 Heaters T 4 A
Fig. 58
Turn the instrument off and disconnect the power plug before replacing the fuses!
Use only the fuse types specified in Chapter "3. Technical Data.
Failure to comply with this can cause severe damage to the instrument and its surroundings
and a risk of fatal injury!
2
1
Fig. 59
• Switchofftheinstrumentatthecircuitbreaker.
• Disconnectthepowerplug.
• Slightlylifttheslidingwindow(1) using the grip (2)
and pull it out to the front.