Leica EG F Heatable forceps Instructions for Use Leica EG F_heatable forceps Order No. 14 0388 83101, RevC 1v3 RevC, English – 11/2012 Always keep this manual near the instrument. Read carefully prior to operating the instrument.
NOTE The information, numerical data, notes and value judgments contained in this manual represent the current state of scientific knowledge and state-of-the-art technology as we understand it following thorough investigation in this field. We are under no obligation to update the present manual periodically and on an ongoing basis according to the latest technical developments, nor to provide our customers with additional copies, updates etc. of this manual.
Table of contents 1. Important Information................................................................................................................ 5 1.1 1.2 1.3 2. Safety............................................................................................................................................ 7 2.1 2.2 3. Safety instructions......................................................................................................................... 7 Warnings........................
1. Important Information 1.1 Symbols in the text and their meanings Symbol for labeling electrical and electronic equipment in accordance with Section 7 of the German Electrical and Electronic Equipment Act (ElektroG). ElektroG is the law regarding the sale, return and environmentally sound disposal of electrical and electronic equipment. Warnings appear in a gray box and are marked by a warning triangle . Notes, i.e.
1. Important Information REF Order number for standard delivery or accessories. SN Designates the serial number of the instrument. The package contents are fragile and must be handled with care. The package must be kept in a dry environment. 1.2 Qualification of personnel • The Leica EG F (heatable forceps) may be operated by trained laboratory personnel only.
2. Safety The safety and caution notes in this chapter must be observed at all times. Be sure to read these instructions, even if you are already familiar with the operation and use of other Leica products. 2.1 Safety instructions These Instructions for Use include important instructions and information related to the operating safety and maintenance of the instrument.
2. Safety 2.2 Warnings To ensure perfect operation of the instrument, the following safety instructions and warnings must be observed: The instrument must be connected to mains only with one of the supplied power cords and only to grounded sockets. Do not interfere with the grounding function by using an extension cord without a ground wire. Extreme temperature fluctuations between storage facility and setup site as well as high humidity may cause condensation to form.
3. Installation 3.1 Unpacking instructions When the instrument is delivered, check the packaging first. Check the shipment for possible damage and note this on the shipping documents if necessary. ò ò ò ð Fig. 3 Fig. 2 • Carefully cut through the adhesive tape at the indicated locations (Fig. 2). • Remove the provided bag with documents (including the User Manual etc.) and replacement fuses. • Then carefully take the control unit (Fig. 4) out of the packaging. • Unpack the forceps (Fig. 4).
3. Installation Necessary assembly work Proceed as follows to make the instrument ready for use: • Forceps holder (1) with paraffin collecting tray (2) is factory-installed on the right side. If it needs to be installed on the left hand side, follow these steps: - Loosen the four Phillips screws (3) on the bottom of the instrument. - Remove forceps holder / collecting tray, reverse it and fasten it to the opposite side (insert screws in corresponding holes on opposite side). 1 3 2 Fig.
3. Installation 3.2 Standard delivery Base unit...............................................................................................................................14 0388 35824 • Control unit with voltage selector switch, 100-120/220-240 V/50-60 Hz • 1 screw-on tweezer holder for installation on the control unit • 1 screw-in heatable forceps with spiral cable • 1 set of fuses (2x T1.25 A L250 V) • 1 set of power cables - 1 power cable for Germany................................................
3. Installation 3.4 Electrical connection The Leica EG F can be connected to different electrical power systems (voltage and frequencydependent, refer to Chapter 4.3 "Technical data"). Please observe the following notes to prevent damage to the instrument: The instrument MUST be connected to a grounded mains socket. Only the power cord intended for the local power supply (socket) may be used.
3. Installation Connecting to mains 1 5 7 8 Before connecting the instrument to the mains, ensure that the voltage selector switch is set to the correct local voltage! An incorrect setting of the voltage selector can cause serious damages to the instrument! The instrument must be disconnected from mains before changing the voltage selection. Connecting the power cord Fig. 7 • Before connecting the power cord, ensure that the mains switch (1) to Off = 0.
4. Instrument Characteristics 4.1 Overview – instrument components/functions Control instrument (front view) Control panel Forceps holder Paraffin collecting tray Heatable forceps (tip width 1 m m ), with cable Fig. 8 Voltage selector Control unit (rear view) Mains switch Socket for power cable Fig. 9 14 Instructions for Use V1.
4. Instrument Characteristics 4.2 Instrument components/functions 1 3 2 The control panel on the front side of the control unit consists of a membrane keypad with push buttons (2), a single-line display (1) and an LED (3) which flashes yellow during the warm-up phase. If the preselected operating temperature is reached, this indicator lights permanently, thereby signaling its readiness.
4. Instrument Characteristics 4.3 Specifications General data Approvals: The instrument-specific approval marks are on the side of the instrument next to the type plate. Nominal voltage (+/- 10 %): 100-120/220-240 V AC Nominal frequency: 50/60 Hz Power fuses: 2x T 1.25 A L250 V Nominal power: 10.
5. Maintenance and Cleaning • Before any maintenance, switch off the instrument and unplug it from power supply. • To clean the exterior surfaces, use a mild and pH-neutral commercial household cleaner.
6. Warranty and Service Warranty Leica Biosystems Nussloch GmbH guarantees that the contractual product delivered has been subjected to a comprehensive quality control procedure based on the Leica in-house testing standards, and that the product is faultless and complies with all technical specifications and/or characteristics warranted.