Manual

29
Leica TP1020
Fig. 12
4. Instrument Setup
4.2 Installation site requirements
The installation site has to fulfill the following
conditions:
• Stableandplaneinstallationsurface.
Minimuminstallationsurfacedimensions:
850 x 850 mm,
• Roomtemperatureconsistentlybetween+5°C
and +40 °C,
• Relativehumidityshouldnotexceed80%.
Necessary conditions for electrical connection
- Grounded wall outlet near enough so it can
be reached with the power cable supplied
together with the instrument.
780 mm
820 mm
Because of the quantity of solvents
used for processing specimens and the
solvent concentrations that build up,
we recommend installing instrument
versions without a fume control system
or active carbon filter in the labora-
tory's fume hood.
The instrument may not be operated in
hazardous locations.
If you wish to install the basket holder
for a second tissue basket, you should
do that now.
For mounting instructions, see Chap-
ter 13 "Optional Accessories".