Manual

30
Instructions for Use V 2.9 RevC, English – 12/2013
4. Initial Operation
4.3 Electrical connection
4.3.1 Checking the voltage selector setting
Fig. 13
The voltage selector is located in a shell (3) above
the power switch (2) at the rear of the instrument.
The actual selected voltage can be seen in the
window (1).
• Checkifthe setting showing inthewindow
(1) corresponds to the nominal voltage in your
laboratory.
• Removetheadhesivetape.
If the setting is correct proceed at 4.3.3.
If the setting does not correspond to the nominal
voltage in your laboratory, the voltage selector
MUST be adjusted to match your laboratory's
electrical setting prior to connecting the instru-
ment to the mains.
230
3
230 V
2
1
Important! The voltage selector is factory preset.
Before connecting the instrument to the power supply, make sure this setting is correct for your
laboratory's electrical connection values.
The socket for connecting the power cable is sealed with an adhesive tape indicating the
factory preset voltage for your particular instrument.
Connecting the instrument to the power with the voltage selector set at a wrong value can
cause severe damage to the instrument!