4. Instrument Setup
4.6.3 Fastening the wax bath connection cables (instruments with fume control system only)
Fig. 19
After having mounted and connected the wax baths, place the magnetic
clamps on the side wall of the housing so that the cables are fixed in a posi-
tion that ensures free movement of the fume containment shields.
Mount the reagent
containers (1) (glass
or aluminum) onto the
station holders (2) of
the platform.
4.6.4 Installing the reagent stations
1
2
Instruments with fume control system are equipped with a Plexiglas
fume containment shield which consists of two parts. To ensure the
fume containment shields can be freely rotated and displaced as re-
quired without being obstructed by the wax bath connection cables,
two magnetic clamps are provided on each of the cables to attach
the cables to the housing.