5. The User Interface
Programming mode buttons
These buttons are used to enter or alter param-
eters when programming.
ARROW LEFT / ARROW RIGHT buttons (cursor)
To move the cursor in the direction of the arrow
to the respectively next data entry position.
• WhenpressingoneoftheARROWbuttons,the
cursor jumps to the next data entry position in
the direction of the arrow.
• ToswitchlinespressARROWLEFT.
PLUS / MINUS buttons
To modify the parameter in the position where the
cursor is located.
• Toincreasethenumericalvalue,pressPLUS.
Todecreasethenumericalvalue,press
MINUS.
• Tochangethenumberofbasketsfrom1to2,
press PLUS.
• Tochangethenumberofbasketsfrom2to1,
press MINUS.
• TochangethestationnumberpressPLUSor
MINUS respectively.
Toswitchonthevacuumfunction(types2and4)
push PLUS.
Todeactivatethevacuumfunction(types2and4),
press MINUS.
These buttons are equipped with a re-
peat function. If one of the ARROW is
pressed for a while, the cursor jumps
to the respectively next data entry
position. If one of the PLUS / MINUS
buttons is pressed for a while, the dis-
played numerical value is continuously
increased or decreased.
The Leica TP1020 is available with and
without a vacuum function.
For the variants without vacuum func-
tion, no "V" appears in the display. Of
course a vacuum function cannot be
activated with the "V" buttons in those
cases.