Manual

52
Instructions for Use V 2.9 RevC, English – 12/2013
7. Preparations for Tissue Processing
7.3 Inserting the tissue basket
Fig. 22
• Fillthetissuecassettesorcapsulesinto
the tissue basket.
• Tohook in the tissue basket(s), the
basket holder(s) should not be located
over a wax bath.
• Liftthecarouselinthemanualprocess-
ing mode.
Rotatethecarouselsothebasket
holder (1) is located above the starting
container for basket 1.
Hookthe tissuebasket(2) into the
basket holder as shown.
• Lowerthebasketintothestartingcon-
tainer in the manual processing mode
or start an automatic processing cycle.
1
2
When working in the manual mode, all carousel movements are activated
through the corresponding function keys on the control panel.
Usecautionwhenloweringthecarousel!Keepyourfingersoutof
the space between the container lid and the upper rim of the con-
tainer!
The carousel may not be rotated manually! Severe damage will
result from doing so!