Manual

60
Instructions for Use V 2.9 RevC, English – 12/2013
Delayed start (continued) • PressPLUS/MINUStoenterthenumberof
days.
• PressARROWtomovethecursortothedata
entry position for the hours.
• PressPLUS/MINUStoenterthenumberof
hours.
• Tomovethecursortothedigitsdestinedfor
entering the minutes, push the ARROW key.
• PressPLUS/MINUStoenterthenumberof
minutes.
The program shown here would be started the
next day at 7.30 PM.
• ToactivatethedelayfunctionpressSTART.
Once the delay function has been activated,
the tissue basket will move immediately to the
selected starting container.
While the basket is waiting in the starting con-
tainer for the actual program to start, the display
reads "- -h - -".
When pressing CLOCK the display switches from
real time indication to reading the selected start-
ing time and the resulting end time.
Display end of processing time
To check if the programmed starting time leads
to an acceptable end of run time:
• Press"CLOCK".
In the example shown here the end of processing
would be the day after next at 9.48 AM.
9. Operating the Instrument in the Automatic Processing Mode
0 1 - - h - -
P 4 1 6 : 3 0