Manual

71
Leica TP1020
10. Warning and Error Codes – Troubleshooting
List of error codes (continued)
Error code "E:11" Loss of process data. - Acknowledge the error message by
pressing any key; "ABORT" message,
the process is canceled
- Restart processing cycle.
- If the same problem recurs, call
Technical Service.
Error code "E:12" Loss of program settings. - Acknowledge the error message by
pressing any key; "ABORT" message,
the process is canceled
- Restart processing cycle.
- If the same problem recurs, call
Technical Service.
Error code "E:13" Erroneous data entry for delayed start
function (e.g. desired starting time
would be earlier than current real time.)
- Correct erroneous data entry.