Manual

81
Leica TP1020
12.3 Replacing the fuses
12. Maintenance
Fig. 25
7
3 1
6
The fuses are located in the voltage selector
shell (3).
• Removetheshell(3) as described in Chap-
ter 4.3.2 "Adjusting the voltage selector".
• Removethefuses(7).
• Insertreplacementfusesofthesametype.
The currently selected voltage can be seen in
the small window (1) in the voltage selector shell.
• Insertthevoltageselectorshelltogetherwith
the fuses back into the receptacle in the instru-
ment and press lightly until it locks.
• Checkifthesettingthatshowsinthewindow
(1) corresponds to the nominal voltage in your
laboratory.
Turn the instrument off and disconnect the power plug before replacing the fuses! Burnt-out
fuses may only be replaced by fuses of the same type and specification.
See Chapter 3 "Technical Data".