12.0 Scanner Power-Up
12.1 Pre-Power-Up Checks
Remove the power cord from the power input receptacle, then turn the power switch to the on
position by pressing the portion of the power switch labeled “I”, as shown in the following figure.
Using a digital multimeter in ohm mode, check the resistance from each point listed in the following
table and indicated in the following figure.
Table 4: Resistance Measurement Points
Point A
Point B
Meter Reading
1
2
Greater than 1M ohm
2
3
Greater than 500K ohm
1
3
Greater than 1M ohm
Figure 94: Power Input Receptacle
Turn the power switch to the off position by pressing the portion of the power switch labeled O”,
and then reattach the power cord to the power input receptacle.