RTR-4 Operator’s Manual Safety Summary
120300 Rev. D SAIC Proprietary 2-3
applicable regulations, and the X-ray Source supplier guidelines,
such as the Golden Engineering XR200 X-Ray Source Operator’s
Manual.
General Safety Procedures
The following are general safety precautions for using X-ray
generating devices:
As Low As Reasonably Achievable (ALARA) principles of
radiation exposure are applicable whenever the RTR-4 system
is being used. Personnel exposure to X-rays shall be kept to the
absolute minimum possible.
Users shall establish appropriate exclusion zones to limit
access to the radiation field. This information can be found in
the Users’s Manual for the applicable X-ray source or on the
manufacturer’s webpage. When two or more exclusion zones
apply, such as using the RTR-4 in the presence (or suspected
presence) of explosives, the larger exclusion zone perimeter
shall apply.
Operating personnel shall review and follow applicable
regulations.
In order to meet the FCC RF safety regulations when using the
wireless configuration, personnel shall maintain a distance of
at least 20cm (8in) from the WiFi antennas when the system
is operational.
Personnel in the vicinity when the RTR-4 system is being
operated may be required to wear an approved X-ray radiation
monitoring device in accordance with local regulations or
administrative procedures.
Personnel shall remain at a safe distance from the X-ray
source, and always remain outside the direct beam during
image acquisition if possible.
Personnel shall remain at a distance of at least 20cm (7.9in)
from the antennas whenever the RTR-4 wireless option is
being used.
When using the optional Notebook Safety Lock, the controller
key shall be turned to the OFF position when not acquiring
images. The X-ray source cannot fire when its controller key is
in the OFF position.
The power key for the X-ray source should be turned to the
OFF position when the system is not acquiring images. The
X-ray Source cannot fire when its key is in the OFF position.