How-To Guide

P51
L10-5 Transducer
Acoustic Output Reporting Table for Track 3 : Transducer ModelL10-5, B-Mode Operation, Probe with 7.1MHz
MI
TIS TIB TI
C
Scan Non-scan
Non-
scan
A
aprt
1
cm
2
A
aprt
> 1
cm
2
Global Maximum Index Value
0.712 0.0105
Associated
Acoustic
Parameter
P
r.a
(MPa) 1.93
P (mW) 1.46 - -
P
α
(z
s
) min and
I
ta,α
( z
s
)min
(m W )
-
z
s
(cm)
z
bp
(cm) -
z
b
(cm) -
z
max of I
pi, α
(cm) 2.279 -
d
ep
(z
b
) (cm) -
f
aw f
(MHz) 7.35 7.35
Dim. of A
aprt
X (cm) 0.44 - - -
Y (cm) 3.84 - - -
Other
Information
t
d
sec) 1334
prr (Hz) 1000
p
r
@I
pi
max
(MPa) 2.888
d
eq
@ I
pi
max
(cm) -
I
pa,α
@MI
max
(W/cm
2
)
Ope
r
a
t
ing
Control
Conditions
Frequency (MHz)
7.1
Number of focus (point)
1
Power (mW)
10
Note
1. No TIS formula information is required for the maximum TIS value in this mode.
2. No information about TIC is required for any transducer assembly that is not used for the head of a transcranial or neonatal
3. If the device meets the 51.2aa) and 51.2dd) exemptions, no MI or TI information is required
(a) The intended use does not include the head, so the TIC is not calculated.
# No data report