Cut Sheet

What is Load Monitoring Inform
Technolog y?
To learn more about Inform
Technology and innovations please visit leviton.com/inform
Specifications
Medical Grade Power Strips
Electrical Specifications
Current 15 A (12 A max continuous load) or 20 A (16 A max continuous load)
Voltage 125 V
Frequency 60 Hz
Surge Technology
MOV
Performance Data
Voltage Protection Rating (VPR)
600 V (L-N), 600 V (L-G), 500 V (N-G)
Max. Continuous Operating Voltage
(MCOV)
150 V
Max. Surge Current, Per Mode
27 kA
Max. Leakage Current <100 μA
Material Specifications
Line Cord #14/3 SJT (15 A models) or #12/3 SJT (20 A models)
Enclosure Powder Coated Steel (18 ga)
Outlet Covers Polypropylene
Mechanical Specifications
HG Plug Configuration NEMA 5-15P or NEMA 5-20P
HG Outlet Configuration NEMA 5-15R or NEMA 5-20R
Environmental Specifications
Operating Temperature 0°C to +40°C
Storage Temperature -10°C to +50°C
Standards
UL/ CSA/ NOM
UL 60601-1/ C22.2 No. 60601-1; UL 60950-1/
C22.2 No. 60950-1; UL 1363A/ C22.2 No. 21; UL 1449/
C22.2 No.8
NFPA NFPA 99 2018 ed
Warranty
Product Warranty 5-Year Limited
Mounting Bracket for Medical Grade Power Strips (5300M-BKT)
Material Specifications
Bracket / Retainer
Powder Coated Aluminum
Mounting Hardware
Stainless Steel
NOM and CSA Standards (C22.2) do not apply to models with Inform™ Technology
Spec is relevant to Surge Protective models only
A simple and ecient way to improve code compliance.
NFPA 99 Article 10.2.3.6 mandates:
The sum of the ampacity of all appliances connected to the outlets does not
exceed 75 percent of the ampacity of the flexible cord supplying the outlets.
Inform
TM
monitors the load of connected devices
so that you don’t have to, eliminating the need
for manual calculations. When the NFPA 99 limit
(75%) has been exceeded, a line-of-sight LED
indication will activate.
OVERCURRENT
INDICATION