User Manual

63
manual. When life-sustaining implants (like pacemakers)
are used along with hearing instruments, Lexie strongly
advises users to always follow the guidelines of the implant
manufacturer as well.
If you suspect any interference to your implant, please
immediately stop using all other electronic devices, including
your hearing instruments, and contact the manufacturer of
the implant as well as Lexie to advise about the suspected
interference.
To ensure our hearing instruments can operate safely and
effectively in close proximity to other electronic devices,
Lexie hearing instruments have been proven to meet the
requirements of the international standard for electromagnetic
compatibility (EMC), IEC 60118-13, and ANSI C63.19.
Life-sustaining implants are required to comply with the
standards described in ANSI/AAMI/ISO 14117 on securing
electromagnetic compatibility with other electronic devices,
including hearing instruments. As a consequence, implants
complying with the ANSI/AAMI/ISO 14117 standards should
continue to operate within the requirements of basic safety and
essential performance when used with hearing instruments
from IntriCon.