51 --How to install..........................................................68 --How to delete.........................................................68 CONNECTION TYPES............. 69 X-ray Generator and Detector Connection of Detector - PC Mode of connecting Network --Trigger Cable Instruction..................................70 --Detector and PC(Wired mode).....................72 --Detector and PC(Wireless mode)................75 CALIBRATION SOFTWARE.... 77 --Calibration Software...................
52 ENGLISH SAFETY INFORMATION Preparation •• Be sure to connect the cables to the proper connectors. Otherwise, the detector may malfunction or may be damaged. •• The power supply provided by LG is designed for the detector from LG. Please contact supply, if any other type of power supply is needed to be used. •• Be sure to fully charge the battery before use. Charge the battery on the day of examination or on the previous day. •• Battery slowly discharges even when not in use.
53 Overall Pressure: 150kg (330.6lb) over the whole area of detector window. Partial Pressure: 100kg (220.4lb) on an area 40mm (1.5 inch) in diameter Disinfection and Cleaning •• Do not spray disinfectants or detergents on the detector. •• When cleaning the detector, be sure to turn off the power, and unplug the power cable from the AC outlet. •• Do not use any flammable chemicals such as thinner, benzene for cleaning. Otherwise, fire or electric shock may result.
54 ENGLISH Overview 14HJ701D is wireless Flat Panel Digital X-ray Detector that can generate images of any part of the body, and designed for a faster approach to digital radiography systems. This model utilize a combination a amorphous silicon TFT glass and high performance scintillator, along with a pixel pitch 127um and 3.9lp/mm of resolution, assure sharp and high contrast image quality. 14HJ701D is X-ray imaging acquisition device that is based on flat-panel.
5 Basic Accessories ENGLISH Detector 1 EA Charger 1 EA 125 Battery 2 EA 90 255 Installation CD 1 EA AC Power adapter 1 EA AC Power cord for Control Box 1EA AC Power cord for AC Power adapter Control Box 1EA Manual 1 EA Main cable 1EA Inspection Report 1EA
56 Optional Accessories ENGLISH Trigger Cable 1EA LAN cable 1EA •• Optional accessories can not be included in accordance with production suffix. CAUTION •• Need to use the authorized components about the below accessories. Unauthorized components may be cause of the damage and malfunction of the product. Component Standard LAN CABLE More than CAT5E Standard Power Cord US – Approved Medical grade regulation Others – Approved country safety regulation The AC/DC adaptors and etc.
57 Detector FRONT ENGLISH PART NAME AND FUNCTION
58 SIDE BACK ENGLISH 1 5 4 2 3 1 1 1 Wireless Antenna 2 Battery Unlock Lever: Unlock lever to remove battery 3 Power Button: Power on/off switch (On : press over 1 sec, Off : press over 5 sec) 4 LED Indicator: Indicating detector’s status 5 Connection to main cable LED Battery Link Power LED Color Green Orange Orange Blinking Green Off Green Green Blinking Off Status Battery is more than 30% charged. Battery charging stauts is 10 ~ 30%. Battery is less than 10 % charged.
59 ENGLISH Battery and Charger 1 Battery: Li-ion polymer battery (Charging time - Typ. 4 Hrs) •• Battery pack itself shows the remaining battery percentage. Battery Remain Indicator Battery Level 75 ~ 100% 50 ~ 75% 25 ~ 50% 0 ~ 25% 2 Battery charger: 3 ports cradle type 3 LED Indicator: Following LEDs are located to each battery - 3 batteries.
60 125 1 2 3 6 5 109.8 ENGLISH Control Box 4 7 1 LED POWER Ethernet Ready Exposure Green Power normal operation Off Power off (AC power cord no connection or Power error.) Green Ethernet normal operation Off Ethernet disconnected. Green Ready signal from X-ray generator is active. Off Ready signal from X-ray generator is inactive. Orange blink Error Orange Exposure signal from X-ray generator is active. Off Exposure signal from X-ray generator is inactive.
61 1 Check the battery mounting hole direction. 3 Press the opposite side to secure the battery indicator. ENGLISH BATTERY INSTALLATION 2 Insert into the hole on the side with the indicator.
62 ENGLISH SPECIFICATION OF EACH PART The product specifications are subject to change without prior notice for product improvements. “ ~ ” refers to alternating current (AC), “ ” refers to direct current (DC). Detector Item Specification Model Sensor type Scintillator Type Total Pixel Matrix Total Pixel Area Pixel Pitch Effective Pixel Matrix 14HJ701D Amorphous Silicon with TFT CsI:TI 3328 x 2816 422.655 (16.63) x 357.632 (14.08) 127 3323 x 2751 A/D Conversion Data Transfer 16 802.
63 NOTE Battery Item Model Rating Size Weight Output Norminal voltage Cycle time Operation Temp Charging time Capacity Battery performance Specification LBQ7222L 7.5 V, 3850 mAh, 30 Wh 204.1 x 10.5 x 7.8 (8.0 x 0.4 x 0.3) Typ. 240 (0.5) 7.5 Max. 800 10 - 35 Typ. 4 Typ. 4000 1400 shots acquired for 3.2 hours (cycle time 8s , without sleep , with Full charged battery) Units mm (inch) g (ib) VDC Cycles ℃ Hours mAh Images Battery Charger Item Specification Model Size LG Battery Charger 125 x 255.
64 ENGLISH Control Box Desc. mm (inch) Model Size Weight Input Output Specification Units LG Control Box LG Control Box 125.0 x 255.0 x 109.8 (4.9 x 10.0 x 4.3) mm (inch) Typ. 1.3 (2.86) kg (Ib) AC Power 100-240 V ~, 50/60 Hz, 1.4-0.7 A AC Inlet Fuse 1 T4L 250V AC Inlet Fuse 2 T4L 250V Power Fuse 1 (F101) T 3.15AH 250V Power Fuse 2 (F102) T 3.15AH 250V 2.1 A, Trigger signals, Ethernet data for Detector A DXD 1 24 V 2.
65 ENGLISH DIMENSION (UNIT: mm/inch) Detector 15 / 0.5 459.5 / 18.0 384 / 15.1 Control Box 125 / 4.9 109.8 / 4.3 255 / 10.
66 7.8 / 0.3 110.5 / 4.3 204.6 / 8.0 Battery Charger 125 / 4.9 90 / 3.5 255 / 10.
67 ENGLISH ENVIRONMENTAL REQUIREMENT PC system requirement PC Specification CPU Memory Disk capacity Network card OS Monitor AP etc intel i5 4 GB At least 10 GB, 500 GB recommended dual ethernet 100/1000 bps Windows 7/8.1/10(32bit, 64bit) Min. Resolution 1280x720 Cisco model is recommended(ex : Linksys EA9200) Environmental Requirements Item Temperature (Storage) Temperature (Operation) Humidity (Storage) Humidity (Operation) Pressure (Storage) Pressure (Operation) Min. -20 10 Typ Max.
68 ENGLISH INSTALLING CALIBRATION SOFTWARE How to install Run the calibration software installation file. Once the installation file has been executed, follow the installation instructions on the screen. How to delete You can delete the Calibration Software in the following ways: Deleting from the Control Panel 1 2 3 4 5 Select Control Panel from the Start menu. Select Programs and Features in Control Panel. Select the LG DXD Calibration on the lists.
69 ENGLISH CONNECTION TYPES X-ray Generator and Detector Select Trigger Mode in accordance with the acquisition method. -- Auto Mode : Detector detects the image obtained after the X-ray -- Manaul Mode : Detector acquires image by pressing generator exposure switch Connection of Detector - PC The connection mode used between the detector and PC. -- Wired Mode: Uses the Control Box to connect the detector to a PC. -- Wireless Mode: Uses a wireless AP to connect the detector to a PC.
70 ENGLISH Trigger Cable Instruction Generator •• Connects a Trigger cable to Port A of the Control Box when there is X-ray generator interface. No.
71 ENGLISH -- EXP_REQ EXP_ACK
72 ENGLISH Detector and PC(Wired mode) Auto Mode Main Cable PC LAN cable Detector Generator Manual Mode Main Cable PC LAN cable Detector Generator Trigger Cable
73 MAKING CONNECTIONS - Wired Connection 1 Open the Network and Sharing Center. -- (Control Panel > Network and Internet > Network and Sharing Center) 2 Right-click on Local Area Connection and enter Properties. ENGLISH 1 Connect the PC to the control box using a LAN cable. Connect the detector to the control box using a link cable.
74 3 Select Internet Protocol Version 4 (TCP/IPv4), then enter Properties to set the following IP address. ENGLISH -- IP address: 10.10.10.2 to 10.10.10.254 (Set it to an IP address other than 10.10.10.100.) -- Subnet mask: 255.255.255.0 -- Gateway: 10.10.10.1 4 Run the LG DXD Calibration. Go to Settings > DXD > Connection&File Save, enter Detector IP (10.10.10.100), then run the Ping to check the connection.
75 ENGLISH Detector and PC(Wireless mode) Auto Mode PC Generator AP LAN cable Detector Manual Mode Generator Trigger Cable Main Cable PC AP LAN cable Detector NOTE Please install the AP and Detector as near as possible without obstacles in between them.
76 MAKING CONNECTIONS-Wireless Connection ENGLISH 1 The Default Wireless Settings as below. -- SSID : LGEDXD -- Password : lgedxd2000 2 It is possible to change the Wireless Settings using LG DXD Calibration Software. •• Please see the Service manual (Wireless AP configuration) for details. 3 Please reboot the Detector after removing the main cable on Detector.
77 To ensure that optimal image quality is obtained from the detector, the detector must be calibrated. The Calibration Software generates and verifies the values necessary for calibration. NOTE •• To ensure optimal image quality, you should calibrate every month for the first three months after the purchase and half-yearly calibration thereafter. •• It is recommended that you leave the detector on for five minutes before calibrating.
78 •• Calibration SW: Configures the settings required for to calibrate software algorithms. ENGLISH •• Firmware Update: Checks the firmware version of the detector or performs the firmware update.
79 Calibration •• Dark and Bright images are obtained from the detector. -- Dark Image: An image obtained without generating X-rays. -- Bright Image: An image obtained by generating X-rays without a phantom or any other object on the detector. •• Generate Avgdark, Offset, and Gain: Used for Corrected Image calculations. -- Corrected Image: An image generated by applying calibration results to a raw image. •• Generate Bad Pixel Map: Bad pixel values are calibrated using surrounding pixel values.
80 User BPM ENGLISH This is used to assign additional bad pixels in BPM.raw after calibrating. Validation This is used to validate the final image by applying calibration results to the X-ray image.
81 Exposure Index ENGLISH This calculates and saves median output value per input dose as a linear expression and a table.
82 ENGLISH Image Functions 4 2 3 1 5 1 Image Viewer •• The Calibration, User BPM, Validation, and EI menus include an image viewer that lets you view the acquired images. •• An image is loaded and displayed when it is acquired or clicked. •• Information about the image is displayed in the following sections. -- Image Loaded: Indicates whether an image is loaded in the image area. * When an image is loaded: Image Loaded. * When no image is loaded: Image not Loaded.
83 2 Navigator 3 Histogram •• Histogram of the acquired image is shown here. •• Window/Level can be adjusted for better image distinction. •• Use the <> buttons or the scroll bars below the Histogram graph to adjust window/level. •• Click the Reset button to restore the default values. 4 Reference Point •• When you click a position inside the Image Viewer, a reference point is specified, and the coordinates and pixel value of the reference point are shown at the top.
84 ENGLISH Log This shows the necessary information about the application execution process that is helpful to the user. It consists of the date, time, type and content, which are saved as a log file.
85 Launching the Program •• Double-click the executable file installed on your PC to launch the calibration software. •• If launching for the first time, the Settings screen appears.
86 ENGLISH Checking IP Address and Ping Test •• A default IP address is entered for the detector. •• If you change the IP address of the detector, enter the new detector IP address in the detector IP field of the calibration tool. •• For instructions on changing the detector IP address, see "Detector and PC(Wired mode)". •• After entering the IP address, set a timeout value and click the Ping button to perform a ping test. A pop-up message appears if the ping test is successful.
87 While the calibration software is run, the acquired images, logs, result files and factory-default calibration results are saved in the designated location. You can change the storage path in the Save location. Click the Apply button when a folder is created in the specified path. NOTE •• Default: C:\Users\heuser\Documents\LG DXD Calibration •• heuser: User name ENGLISH Checking Save Location
88 ENGLISH Apply After performing the ping test and verifying the save location, click the Apply button to perform the following procedures 1 Automatically create the necessary folders in the save location designated in the Checking Save Location stage. 2 Load from the detector and save the factory-default calibration results. 3 Load the detector's settings.
89 Upon Apply, the current detector settings are loaded on the settings screen below. •• Detector Parameters are parameters used when obtaining images from the detector. •• Click the Save button to apply the settings entered. •• Descriptions of the data are as follows: -- Trigger Mode : This sets Trigger Mode. * Auto : Enables the Auto Exposure Detection feature. * Manual : Disables the Auto Exposure Detection feature. -- Sensitivity : This is the panel sensitivity.
90 ENGLISH Checking and Changing Calibration Software Settings The calibration parameters are updated when the Calibration Software tab is clicked. Calibration Parameters These settings are used to perform calibration. •• Descriptions of the data are as follows: -- Target Gain : Amplification level within algorithms. Pixels with values that exceed the Gain Margin are specified as bad pixels. -- Offset Margin : Offset margin value used within algorithms.
91 •• Rotation : This sets the rotation angle of the image. (0°, 90°,180°, 270°) •• Flip : Sets how the image shown in the Image Viewer is rotated. (None, Horizontal, Vertical) •• Invert : Inverts the image data shown in the Image Viewer. •• Click the Save button to apply the settings entered. •• Descriptions of the buttons are as follows. -- Save : Applies the changes. -- Reset : Loads the factory default settings. -- Cancel : Loads the last saved values. -- Exit : Return to the previous screen.
92 ENGLISH Calibration After configuring all settings, click the Calibration tab to open the Calibration menu. NOTE •• This tab cannot be accessed until the settings have been configured. 1 5 2 3 4 1 Open Calibration Menu -- Click the Calibration menu button to open the menu. 2 Get Dark The menu is used to acquire dark images necessary for performing calibration.
93 NOTE •• All the images are deleted when calibration ends. For backup, copy the images and paste them into another folder. •• Up to 10 images can be saved. •• Bright Image: An image obtained by generating X-rays without a phantom or any other object on the detector. 4 Calibration This area is used for calibration. •• At least four dark images and four bright images are required for calibration.
94 ENGLISH User BPM The user can manually edit the Bad Pixel Map created during Calibration in this menu. NOTE •• You can skip the User BPM stage and proceed to Validation. 1 2 3 4 1 Open User BPM Menu -- Click the User BPM menu button to open the menu. NOTE •• The pop-up window below will appear if you open this menu without performing calibration. 2 Check History File -- Check that the History name created during calibration matches the name shown in the current History window.
95 3 Image Acquisition 4 Pixel View •• Check pixel values in Pixel View -- Pixel values for the center of the Image Viewer are displayed in the Pixel View window. -- The minimum value, maximum value and pixels estimated as bad pixels are shown as below. * Minimum value : Indicated as a green number * Maximum value : Indicated as a red number * Estimated bad pixels : Indicated with gray background -- The size of the Pixel View window can be adjusted using the icon.
96 ENGLISH Validation This procedure allows you to perform a visual inspection of the calibration results after completing calibration. 1 2 3 4 1 Open Validation Menu -- Click the Validation menu button to open the menu. NOTE •• The pop-up window below will appear if you open this menu without performing calibration. 2 Check History File -- Please confirm whether the name of the History file created during calibration matches the name displayed in the History window.
97 3 Image Acquisition (Procedure matches that from the User BPM stage) 4 Apply/Unapply Calibration Results -- You can apply/unapply calibration results (Dark Average Map, Offset Map, Gain Map, Bad Pixel Map) to the acquired image. : Apply : Unapply NOTE •• When an image is acquired and loaded for the first time, all results are applied by default. •• If no image has been acquired, the Apply/Unapply button is disabled.
98 ENGLISH EI (Exposure Index) 1 2 3 4 5 1 Open EI Menu -- Click the EI menu button to open the menu. NOTE •• The pop-up window below will appear if you open this menu without performing calibration. 2 Check History File -- Please confirm whether the name of the History file created during calibration matches the name displayed in the History window.
99 4 Enter Dose Value 5 Measure & Save -- When you click the Measure & Save button after acquiring an image and entering a dose value, the popup notification below appears. -- The results are saved in the save location specified for EI calculation. (Example: C:\Users\heuser\Documents\LG DXD Calibration\Serial Number\Calibration Results Folder (Time and Date) heuser: Username) NOTE •• When Measure & Save is repeated, the results file in the folder is updated.
100 ENGLISH Exit When you click the Quit button, the calibration tool quit popup window appears. In the popup window below, click the Yes button to quit or click the No button to return to the previous screen.
101 When you click the About button in Settings, the following popup window appears. It shows the information about the application.
102 ENGLISH General PopUp The following are descriptions of the general pop-up windows that may appear when the calibration software is being used. Image Acquisition Cancellation •• When the Acquisition button is clicked for image Acquisition, it will be replaced with a Cancel button while the image is being acquired. •• Once image acquisition is complete, the button will revert to an Acquisition button. •• During image acquisition, you can cancel the process by clicking the Cancel button.
103 Image Acquisition Failure ENGLISH •• The popup window below appears when image acquisition fails. Check the network and Detector status, then try again.
104 ENGLISH SERVICE MANUAL Detector Wired IP Address Setting 1 From the Usage section, perform "Launching the Program", "Checking IP Address and Ping Test", "Checking Save Location" and "Apply". 2 Click the Change DXD IP button. 3 At the popup window, change the setting and click the Apply button. -- When the Apply button is clicked, the system attempts to change the IP address.
105 4 Check the result and restart the detector. -- Once the new IP address is applied, restart the detector so that the new setting takes effect. -- Restart the detector automatically using the program. -- Because the detector is disconnected when restarting, perform the Connection & File Save process again. ENGLISH -- The following popup windows appear, depending on the successful and failed.
106 ENGLISH Wireless AP configuration To have the Detector connect wirelessly to an AP, the AP information must be saved in the Detector in advance. Once the AP information is transferred and saved in the Detector, the Detector attempts to connect to the AP. The saved AP information can be viewed in the Web Monitoring feature. 1 From the Usage section, perform "Launching the Program", "Checking IP Address and Ping Test", "Checking Save Location" and "Apply".
107 3 Check results. ENGLISH -- The following popup windows appear, depending on the result.
108 ENGLISH Detector Firmware Update This menu allows you to check the firmware version of the detector or update the firmware. 1 From the Usage section, perform "Launching the Program", "Checking IP Address and Ping Test", "Checking Save Location" and "Apply". 2 Select the Firmware Update tab. 2 3 4 5 3 Check the current firmware version. -- The current firmware version of the detector is shown here when the PC is connected to the detector. 4 Select a firmware file to update.
109 5 Perform the update. CAUTION •• Do not remove the power cable until the update is complete. The update may not be successful if the power is disconnected during the update process. 6 Check results. -- When the update is complete, the result is shown in a popup window. ENGLISH -- Select a file, then click the Update button to start the Update process.
110 ENGLISH Storage of installation date The date of the first calibration can be saved in the detector. 1 From the Usage section, perform "Launching the Program", "Checking IP Address and Ping Test", "Checking Save Location" and "Apply". 2 Select the Settings > DXD tab. 3 Check the installation date and select a date format. -- YYYY: Year -- MM: Month -- DD: Day NOTE •• The date is displayed based on the OS information. 4 Click the Register button to open a pop-up window.
111 This menu allows you to save the Power Options of detector. 1 From the Usage section, perform "Launching the Program", "Checking IP Address and Ping Test", "Checking Save Location" and "Apply". 2 Select the Settings > DXD tab. 3 Select the options of Auto Sleep and Auto Power-Off. 4 Click the Save button to save the options on Detector. NOTE •• Calibration Software only saves the options. •• Detector enters the the Power Mode when there is no communication during the setting times.
112 ENGLISH Web monitoring This feature displays the detector's release date, installation date, software version and other internally managed information in a web browser. Internal Information Category Product information Network Battery ETC Content Software version Product release date Product installation date Model number Serial No.
113 1 Establish a wired/wireless connection between the detector and the PC. -- Refer to the "Detector and PC(Wireless mode)" or "Detector and PC(Wireless mode)" sections for instructions. 2 Enter the IP address in the address input of the detector PC web browser. 3 The default IP address: 10.10.10.100 enter the page as below appears.
114 ENGLISH MAINTENANCE Cleaning •• Make sure to turn the detector off and remove the battery before cleaning. Inspection •• Check the detector regularly to ensure that it operates stably and properly. If a problem occurs, please contact the manufacture. •• Use the checklist below as a reference for checking the product.
115 Problems that occur while using the detector can be solved by using the information on this page. If problems persist, please contact the manufacturer. Problem The product does not turn on. The product was suddenly turned off during use. The Control Box Ready/Exposure LED flashes in orange. The PC and the detector are not connected. There is a problem with acquired images. Action •• Check that the power cord is properly connected.
116 ENGLISH SOLUTION FOR FIREWALL BLOCK If Calibration Software connection doesn’t work for firewall in Windows 8, refer to instructions as below. 1 Enter into control panel and click on the System and Security link 2 Select the Windows Firewall link. 3 Click on the Advanced settings link situated in the pane on the left hand side.
117 4 Choose Inbound rules in Windows Firewall with Advanced Security. Scroll and find File and Printer Sharing rule 5 Click Enable Rule. 6 Check its status and try to connect detector again. ENGLISH to enable.
118 ENGLISH SUPPLEMENT. WIRELESS AP SET UP INSTRUCTION (MODEL : Cisco Linksys EA9200) 1 Connect the LAN Cable from the Ethernet port on the PC to the Ethernet port on the AP. 2 Launch your web browser and enter linksyssmartwifi.com or http://192.168.1.1 in the Address bar then press Enter (IP number address for the 1st access is 192.168.1.1. However, IP number address for accessing will be 10.10.10.1 after changing 10.10.10.1) Enter into Connectivity > Local Network.
119 3 Enter into Wireless. You can change network name and password like below. ENGLISH (You should click Apply button to apply current setting) For more information, please visit the web site as below. http://www.linksys.
Model Serial No. To obtain the source code under GPL, LGPL, MPL, and other open source licenses, that is contained in this product, please visit http://opensource.lge.com. In addition to the source code, all referred license terms, warranty disclaimers and copyright notices are available for download.
이 제품은 국내(대한민국)용입니다. 전원(전압, 주파수)이 다른 해외에서는 사용할 수 없습니다. 본 제품에 탑재된 GPL, LGPL, MPL 및 그 외 오픈소스 라이선스 하에 개발된 소스 코드를 얻으려면, http://opensource.lge.com를 방문하십시오. 소스 코드를 비롯하여 해당하는 모든 라이선스의 내용 및 저작권 표시, 보증 책임이 없다는 표시를 다운로드할 수 있습니다. 본 제품을 구매하신 후 3 년 내에 opensource@lge.com 로 e-mail을 보내 본 제품에 탑재된 소프트웨어의 오픈 소스 코드를 요청하실 경우, 매체 비용, 운반비 등 제공에 필요한 최소한의 비용을 받고 CD-ROM에 담아 제공해 드립니다.
이 제품은 국내(대한민국)용입니다. 전원(전압, 주파수)이 다른 해외에서는 사용할 수 없습니다. 본 제품에 탑재된 GPL, LGPL, MPL 및 그 외 오픈소스 라이선스 하에 개발된 소스 코드를 얻으려면, http://opensource.lge.com를 방문하십시오. 소스 코드를 비롯하여 해당하는 모든 라이선스의 내용 및 저작권 표시, 보증 책임이 없다는 표시를 다운로드할 수 있습니다. 본 제품을 구매하신 후 3 년 내에 opensource@lge.com 로 e-mail을 보내 본 제품에 탑재된 소프트웨어의 오픈 소스 코드를 요청하실 경우, 매체 비용, 운반비 등 제공에 필요한 최소한의 비용을 받고 CD-ROM에 담아 제공해 드립니다.