Use Instructions

57
Questions? Call 1-800-635-6156
Table 2 – Immunity
The GoSafe 2 Mobile Pendant (7250MHB) complies with the immunity standards and
levels specied in the table below.
Immunity Test Compliance Level
Electrostatic Discharge (ESD) EN/IEC
61000-4-2
±8 kV contact discharge
±2, 4, 6, 8, 15 kV air discharge
Electrical fast transient/burst EN/IEC
61000-4-4
(only for battery charger power supply)
±2 kV for power supply lines
Surge EN/IEC 61000-4-5
(only for battery charger power supply)
±1 kV line(s) to line(s)
Power Frequency 50/60Hz Magnetic
Field EN/IEC 61000-4-8
30A/m
Voltage dips,
short interruptions
and voltage variations on power supply
input lines EN/IEC 61000-4-11
(only for battery charger power supply)
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0 % UT; 1 cycle & 70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
(voltage interruption)
Conducted RF EN/IEC 61000-4-6 6 Vrms
Radiated RF EN/IEC 61000-4-3 10 V/m
80 MHz - 2.7 GHz
80 % AM 1 kHz
and
Proximity Fields;
9 V/m to 28 V/m15 Spot frequencies
(MHz): 385, 450, 710, 745, 780, 810, 870,
930, 1720, 1845, 1970, 2450, 5240, 5500,
and 5785 MHz
NOTE: Conducted RF Immunity does not apply to an internal battery operated device
(7250MHB). Limit above is set for the 7250MHB battery charger only.