LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 1 ® ® BLOOD GLUCOSE MONITORING SYSTEM SYSTÈME DE CONTRÔLE DE LA GLYCÉMIE OWNER’S BOOKLET MANUEL DU PROPRIÉTAIRE
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 3 ® ® BLOOD GLUCOSE MONITORING SYSTEM OWNER’S BOOKLET
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 4 Dear ONE TOUCH® BASIC® Owner: Congratulations! You’ve chosen a very easyto-use blood glucose monitoring system for home testing. Your ONE TOUCH® BASIC® System will give you the accurate test results you need, in just a few simple steps. Everything you need to know about using the ONE TOUCH BASIC System is included in this booklet. Blood glucose monitoring plays an important role in controlling your diabetes.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 5 CAUTION: Before using any product to test your blood glucose (sugar), read all instructions and practice the test. Consult your diabetes health professional for advice regarding selection of equipment, testing times and procedures. These recommendations apply to all blood glucose monitoring systems and are supported by the Diabetes Educator Section of the Canadian Diabetes Association.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 6 TABLE OF CONTENTS 1. GETTING STARTED Description and Use . . . . . . . . . . . . . . . . 1 Getting to Know the Meter . . . . . . . . . . . 2 Precautions and Limitations . . . . . . . . . . 5 2. SETTING THE METER Selecting the Display Language and the Unit of Measurement . . . . . . . 7 Coding the Meter . . . . . . . . . . . . . . . . . . 8 3. CHECKING THE SYSTEM Checking with the Check Strip . . . . . . . 11 Checking with Control Solution . . . . . . 17 4.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 7 6. AFTER TESTING Checking the Amount of Blood . . . . . . 36 Recording Your Test Results . . . . . . . . . 37 Displaying Last Test Result . . . . . . . . . 38 7. HELP! (Troubleshooting) Test Procedure Messages . . . . . . . . . . . 40 Error Messages . . . . . . . . . . . . . . . . . . . 43 8. TAKING CARE OF THE METER Daily Meter Check . . . . . . . . . . . . . . . . 55 Cleaning the Meter . . . . . . . . . . . . . . . . 55 Battery Replacement . . . . . . . .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 8 BLANK
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 1 1. GETTING STARTED Description and Use The ONE TOUCH® BASIC® Blood Glucose Monitoring System is intended for in vitro diagnostic use. This statement means that the system should be used only for testing purposes and only outside of the human body. The ONE TOUCH® BASIC® Meter and Test Strips are used for monitoring whole blood glucose for people who have been diagnosed with diabetes mellitus. This device is not to be used for the diagnosis of diabetes.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 2 Getting to Know the Meter Study this diagram and become familiar with all the parts of your ONE TOUCH® BASIC® Meter.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 3 ON/OFF BUTTON. This button turns the Meter on and off. Button to select display languages in English, French, Spanish, German, Dutch, Italian, Portuguese, Swedish, Danish, Norwegian, Finnish, Greek, Turkish, Hungarian, Czech, Polish, Russian or Japanese and each language’s corresponding unit of measurement, (mmol/L) or (mg/dL). DISPLAY. This is where you read your test results and the simple messages that help guide you through the test. BATTERY COMPARTMENT.
Page 4 TEST AREA. When the Test Strip Holder is removed, the Test Area is exposed. The Test Area has a clear, protective coating over the Meter optics window that must be kept clean and lint-free for accurate results. Be careful not to scratch or damage this coating. to use them. DO NOT store Test Strips outside of their packaging.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 5 Precautions and Limitations Never make significant changes to your medication program or ignore physical symptoms without consulting your Health Care Professional. The following information may be useful to you and your Health Care Professional when using the ONE TOUCH® BASIC® System to monitor your blood glucose level.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 6 DO NOT use ONE TOUCH® Test Strips beyond the expiration date printed on the package label. If using Test Strips from a vial, discard the vial and unused Test Strips 4 months after the date first opened. DO NOT store or carry Test Strips outside of their vial or foil package. ONE TOUCH Test Strips must be stored in a cool, dry place, below 30°C (86°F). Do not refrigerate Test Strips or place Test Strips in heat or direct sunlight.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 7 2. SETTING THE METER 1. Press and hold the “C” Button. 2. Press and release the On/Off Button. 3. Release the “C” Button. Selecting the Display Language and the Unit of Measurement The ONE TOUCH® BASIC® Meter displays easy-to-follow messages in English, French, Spanish, German, Dutch, Italian, Portuguese, Swedish, Danish, Norwegian, Finnish, Greek, Turkish, Hungarian, Czech, Polish, Russian, or Japanese to help guide you through the test procedure.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 8 Norwegian (mmol/L), “P 15” for Finnish (mmol/L), “P 16” for Greek (mg/dL), “P 17” for Turkish (mg/dL), “P 18” for Hungarian (mmol/L), “P 19” for Czech (mmol/L), “P 20” for Polish (mmol/L), “P 21” for Russian (mmol/L), “P 22” for Japanese (mg/dL), “P 23” for Japanese (mmol/L), “P 1” for English (mg/dL), “P 2” for Spanish (mg/dL), “P 3” for English (mmol/L).
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 9 Step 1: Turn on the Meter Then Press the On/Off Button to turn the Meter on. The result from your last blood glucose or Control Solution test will appear for 3 seconds. memory or mm 5 . 9 (example) code 7 code 7 (example) appears on the display for 2 seconds, followed by memory c 5 .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 10 Step 2: Match the Code Number The Meter is now properly coded and you may proceed with the test. From now on, the Meter will remember this code until you change the code for a new package of Test Strips. Press the “C” Button. With code 7 (example) on the display, press and release the “C” Button, and the number will increase by one. Continue pressing until the number on the display matches the code number on your Test Strip package.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 11 Checking with the Check Strip 3. CHECKING THE SYSTEM A purple and white Check Strip is included with your ONE TOUCH BASIC Meter. The Check Strip is used to make sure your Meter is working properly. There are two ways to make sure your ONE TOUCH® BASIC® System is working properly. The Check Strip is used to check that the Meter is operating properly.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 12 Before doing a Check Strip test, make sure the Test Strip Holder, the Test Area and Check Strip are clean, dry, and lint-free. Do the Check Strip test at room temperature between 18°C and 26°C (64°F–79°F). then code 7 will appear on the display. How to use the Check Strip Step 1: Press On/Off Button. Your last blood glucose or Control Solution test result will appear for 3 seconds, memory or mm 5 . 9 (example) (example) memory c 5 .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 13 When Step 2: insert strip When strip appears, slide the notched end of the Check Strip into the Test Strip Holder with Side 1 (purple) facing up. apply sample appears, slide the Check Strip from the Test Strip Holder. 1 ✔ wait appears for a few seconds.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 14 Step 3: appears if the number falls within the acceptable Check Strip range. The Meter automatically tells you if your reading falls within the correct Check Strip range. (The range is also printed on the back of the Meter.) When side 2 insert side 2 appears, turn the Check Strip over and slide it back into the Test Strip Holder, notched end first, Side 2 (white) facing up. The Meter counts down from 4 to 0, then A 4 .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 15 If the reading is not within the printed range, the Meter displays will read glucose or Control Solution test result will appear for 3 seconds the next time you turn on your Meter, then A 2. 1 (example) Anotok redo A redo A code 7 Go back to Step 1 and repeat the Check Strip test. If this message appears again, clean your Meter and repeat the test. For further assistance, call LifeScan Technical Services.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 16 ■ CAUTION: If your Check Strip result falls outside the printed range, the Meter is not working properly. DO NOT use the Meter to test your blood until you get a Check Strip reading that is within the printed range. For assistance, call LifeScan Technical Services. A 2. 1 (example) Anotok redo A Cleaning and Maintaining the Check Strip ■ You must use the Check Strip: ■ ■ ■ Whenever this message appears: At least once a day. After cleaning the Meter.
LSI-97-050 OTB Eg ■ 7/25/97 11:09 PM Page 17 If you lose or damage the Check Strip, call LifeScan Technical Services for a free replacement. ONE TOUCH® Normal Glucose Control Solution—Blue Formula should be used to make sure your Meter and Test Strips are working properly. Checking with Control Solution Normal Glucose Control Solution - Blue Formula R Before you use the Meter to test your blood glucose for the first time, practice the procedure using the Control Solution in place of blood.
LSI-97-050 OTB Eg ■ ■ ■ 7/25/97 11:09 PM Page 18 Use only ONE TOUCH® Normal Glucose Control Solution—Blue Formula, available from your drug store or Authorized LifeScan Distributor. Check the expiration date on the Control Solution vial. If it has expired, replace it with new Control Solution. Shake the Control Solution vial vigorously before using. To do a Control Solution test, follow the same procedure you would if you were testing your blood. (See Testing Your Blood, page 27.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 19 The range for the ONE TOUCH® Normal Glucose Control Solution— Blue Formula is printed on the Test Strip vial or the foil wrapper. This range is for the Glucose Control Solution only; it is not intended as a recommended range for your blood glucose test results. You should use the Glucose Control Solution: ■ ■ ■ CAUTION: If your Control Solution result falls outside the printed range, the system is not working properly.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 20 Using the PENLET II Automatic Blood Sampler 4. GETTING A DROP OF BLOOD The PENLET® II Automatic Blood Sampler is an easy, safe tool for obtaining a good drop of blood for testing. Sliding Barrel Cap Step 1: Insert a Lancet Release Button Remove the PENLET II Cap by pulling it straight off.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 21 Insert a new, sterile Lancet into the Lancet Holder. The Lancet will slide into the Lancet Holder easier if you DO NOT line up the ridges on the Lancet with the slots in the Lancet Holder. (NOTE: Inserting the Lancet may automatically cock the PENLET ® II.) Hold the Lancet firmly and gently twist off the Lancet Protective Disk.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 22 Regular Deep Penetration Replace the PENLET® II Cap. The PENLET II Sampler includes two Caps. The Cap that comes attached to the PENLET II Sampler has a single line on the flat side and works well for children and most adults. The other Cap has two lines on the flat side and works well for very thick or calloused skin, or when a deeper puncture is needed.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 23 Step 2: Cock the PENLET ® II Sampler Step 3: Get a Drop of Blood Wash your hands with soap and warm water and dry them thoroughly. Warm water stimulates the flow of blood to the fingers. Hanging your arm down at your side for 10–15 seconds before the fingerstick will make it easier, too. If you use alcohol to clean your finger, make sure you let it dry before sticking your finger.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 24 Choose a spot on the side of a different finger each time you test. Repeated punctures in the same spot can make your finger sore and calloused. Hold the PENLET® II Sampler firmly against the side of the finger, with the Cap resting on the finger. (The harder you press, the deeper the puncture.) Press the dark gray Release Button.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 25 Step 4: Remove the Lancet Always use caution when removing the Lancet and PENLET ® II Cap. Remove the PENLET II Cap. Grasp the ridges on the dark gray prongs located on the shaft near the Release Button. Point the Lancet down and away from you. Pull back on the dark gray sliding barrel until the Lancet drops out. Squeeze the finger gently to obtain a large, hanging drop of blood. Proceed with the blood glucose test. (See Testing Your Blood, page 27.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 26 Cleaning the PENLET II Sampler CAUTION: ■ Never use a Lancet that has been used by someone else. ■ Dispose of the Lancet in a container for sharp objects. ■ To help avoid infection, use a new sterile Lancet every time you test. ■ If you share a PENLET® II Sampler, each person should always use a new Lancet and a new or properly disinfected Cap. For assistance, call LifeScan Technical Services. Clean the PENLET II Sampler and Cap with soap and water as needed.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 27 5. TESTING YOUR BLOOD You can test your blood glucose by following these three simple steps: insert mm 6. 0 PAS D'ALCOOL Step 1: Press On/Off Button (last test result appears), insert Test Strip. Step 2: Apply blood sample. Step 3: Accurate results in 45 seconds. NOTE: Be sure to read the following section carefully before testing.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 28 Step-by-Step Instructions IMPORTANT: ■ The Test Strip Holder and Test Area of the Meter must be clean, dry and lint-free. ■ Check the expiration date on the Test Strip package. If the date has passed, discard the Test Strips and open a new package. ■ If using vial Test Strips, discard unused Test Strips 4 months after opening. ■ If you have questions about the testing procedure, call LifeScan Technical Services. Choose a clean, dry work surface.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 29 T E S T S T R I P R CODE: 10 OPEN HERE TEST FOR GLUCOSE IN WHOLE BLOOD INC. Milpites,CA Made in U.S.A.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 30 Step 1: Press On/ Off Button to turn the Meter on Then code 10 • appears on the display for 2 seconds. Make sure the code number on the Meter display matches the code number on the Test Strip package you are using. (See Coding the Meter, page 8.) Your last blood glucose or Control Solution test result will appear for 3 seconds. memory or mm 5 . 9 (example) (example) The message insert strip memory c 5 . 4 (example) then appears.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 31 Insert Test Strip T R 53B .A. 51-4 INC. e in U.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 32 The Test Spot should be white or ivory-colored, with no tears or wrinkles. insert Test Spot When insert strip wait appears, slide the notched end of the Test Strip into the Holder with the Test Spot side up. Make sure you push the Test Strip all the way into the Holder until it stops. appears for a few seconds. Then, apply sample appears until you apply blood to the Test Spot (or for 5 minutes).
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 33 Step 2: Apply blood sample Make sure that you: With the Test Strip in the Meter and ■ apply sample ■ on the display, apply a drop of blood to the Test Spot. ■ sample ■ ■ The Meter will beep when the blood has been applied to the Test Spot. 33 Do not remove the Test Strip from the Meter to apply blood. Do not smear blood on the Test Spot or apply a second drop after the test begins. Do not move the Test Strip when you are applying blood.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 34 Step 3: Accurate results in 45 seconds Be sure to wait for the beeps before you note your mm 6. 0 result. Your result will continue to display until you turn the meter off or for 2 minutes, when the meter automatically turns off. The Meter displays results which are between 0 and 33.3 mmol/L. Results above 33.3 mmol/L are displayed as The Meter then counts down from 45 to 0 seconds, followed by a series of beeps when your result is displayed. mm 6 .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 35 6. AFTER TESTING c 5 . 8 (example) contrl Although the ONE TOUCH® BASIC® System requires only a small drop of blood, it is very important that the drop be large enough to cover the Test Spot completely. appears on the display, your drop of blood was too small or smeared, or the Test Strip was inserted only partway into the Test Strip Holder. Repeat the test with a new Test Strip and enough blood to cover the entire Test Spot.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 36 Checking the Amount of Blood on the Test Strip The Meter cannot always tell if a sample is too small, so it is important that you also look at the Test Strip to make sure that you applied enough blood. Front ■ ■ Side Remove the Test Strip from the Meter and look at the Test Spot. It should still have a wet, shiny drop that completely covers the circle.
LSI-97-050 OTB Eg 7/25/97 11:09 PM (Incorrect) Page 37 If there are any white patches or streaks, you may have smeared the blood or the drop was too small. In either case, you may have an inaccurately low result. Repeat the test with a new Test Strip. Recording Your Test Results Your ONE TOUCH® BASIC® System Kit comes with a Logbook that you can use to keep a permanent record of your test results, along with information on diet and medication.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 38 Only blood glucose or Control Solution test results are recalled and displayed. Check Strip test results are not stored in the Meter memory. Displaying Last Test Result The ONE TOUCH® BASIC® Meter automatically recalls and displays your last result—either from a blood glucose or Control Solution test—whenever the Meter is turned on. memory or mm 5 . 9 (example) When using the Meter for the first time memory nodata memory c 5 .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 39 7. HELP! (Troubleshooting) The display messages that appear on the Meter will guide you through the glucose test and quality control test procedures and alert you to problems as they occur. This section provides you with a complete listing of the display messages, what they mean, and what to do if there is a problem.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 40 Test Procedure Messages These messages appear during routine testing. Just follow them carefully and they will guide you through the test. MESSAGE APPEARS WHEN You press and hold the On/Off Button. It provides a visual test that all display segments are working. memory mm 6 . 0 (example) The Meter is turned on. The value displayed is your last blood glucose test result. (When using the Meter for the first time, memory nodata will appear.) memory c 5 .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 41 MESSAGE APPEARS WHEN insert strip The Meter is ready for you to perform a blood, Check Strip, or Control Solution test. wait The Meter is performing internal checks. apply sample The Meter is ready to receive a blood or Control Solution sample. If performing a Check Strip test, this is your signal to remove the Check Strip (Side 1) from the Meter. 45, 44, 43...0 The Meter is counting down from 45 seconds to 0. mm 6 . 0 A blood test is completed.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 42 MESSAGE APPEARS WHEN c 5 . 4 (example) contrl A Control Solution test is completed. However, if this appears after a blood test, it means that your test result was not accurate because the blood sample was too small and the Meter read it as a Control Solution sample. insert side 2 You are performing a Check Strip test. Turn the Check Strip over and slide it back into the Test Strip Holder with Side 2 up. A 4 .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 43 MESSAGE APPEARS WHEN p 1R23 You are setting the display language and the unit of measurement. This means that messages will appear in the selection program. “P 3” corresponds to the English (mmol/L) selection. “P 4” corresponds to the French (mmol/L) selection. Error Messages When any of these messages appear, there is a problem with the Meter or the way in which you are performing a procedure. In most cases, problems are easy to fix.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 44 MESSAGE PROBLEM WHAT TO DO 114 The language and unit of measurement have been changed to “P 1” English (mg/dL) either accidentally or through an electronic malfunction. See Setting the Meter, page 7, to reset the language and unit of measurement to “P 3” English (mmol/L). battry The battery is too low. The Meter will not operate. Replace the battery. clean test area 1) There is dirt, blood, or lint on the Test Area.
LSI-97-050 OTB Eg 7/25/97 11:09 PM MESSAGE c 3 . 8 (example) contrl Page 45 PROBLEM WHAT TO DO 3) The Test Strip was inserted before insert strip appeared on the display. Repeat the test. Wait for insert strip before inserting a new Test Strip. 4) The center of the Test Strip Holder where the small hole is located is raised. Depress the center by pressing forward on the base, or by pressing down on the center, or by inserting a check strip into the Test Strip Holder.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 46 MESSAGE PROBLEM WHAT TO DO error1 retest The sample was applied before apply sample appeared. Repeat test with a new Test Strip and wait for apply sample to appear. error2 retest Error in the test procedure such as: 1) The Test Strip moved during the test. Repeat the test with a new Test Strip. 2) The Test Strip was not inserted correctly. Repeat the test, pushing a new Test Strip all the way into the Test Strip Holder.
LSI-97-050 OTB Eg MESSAGE 7/25/97 11:09 PM Page 47 PROBLEM WHAT TO DO 4) There was not enough blood on the Test Strip. Repeat the test with a new Test Strip and a larger drop of blood. 5) The Meter was used in very bright light. Move the Meter out of direct light and repeat the test with a new Test Strip. 6) The Check Strip procedure was incorrect. Repeat the Check Strip test. 7) The Meter may not be operating correctly. Call LifeScan Technical Services.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 48 MESSAGE PROBLEM WHAT TO DO insert insert insert When any message flashes continually, it means the battery is getting low. Test results will still be accurate, but replace the battery as soon as possible.
LSI-97-050 OTB Eg MESSAGE 7/25/97 11:09 PM Page 49 PROBLEM WHAT TO DO 2) An error2 retest message was displayed during your last test due to a procedural error. Review the error2 retest problems listed on pages 46–47 before repeating the test. 3) A not ok message was displayed during your last test. Call LifeScan Technical Services. 4) A power loss, including turning the meter off, occurred during your last test. Repeat the test with a new Test Strip.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 50 MESSAGE PROBLEM WHAT TO DO not enough blood retest 1) The blood sample was too small or smeared. Repeat the test with a new Test Strip and a large, shiny drop of blood. 2) The Test Strip was not inserted all the way into the Test Strip Holder. Repeat the test. This time push the Test Strip all the way into the Test Strip Holder. not ok Your Meter may have an electronic problem. Call LifeScan Technical Services. A 5 .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 51 MESSAGE PROBLEM WHAT TO DO redo A code 7 The last Check Strip test was outside the acceptable range and a repeat test that passed has not been performed. Stop and do a Check Strip test. The Test Strip was inserted before the Meter was ready. Remove the Test Strip. Wait for insert strip message.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 52 MESSAGE PROBLEM WHAT TO DO reset code The Meter has lost some important information, including the code number, and the language and the unit of measurement have changed to “P 1” English (mg/dL) either accidentally or through an electronic malfunction. Use the “C” Button to reset the Meter code to match the Test Strip package code. Reset the language and unit of measurement to “P 3” for English (mmol/L) as described in the section Setting the Meter, page 7.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 53 8. TAKING CARE OF THE METER Strip Holder is still securely in place and check the Meter with the Check Strip before doing a blood glucose test. Your ONE TOUCH® BASIC® Meter is simple to use and easy to maintain. However, it must be handled carefully and cleaned regularly to remain in good operating condition. You should follow these rules at all times: ■ Keep the Test Strip Holder and Test Area clean. ■ Do not drop or toss the Meter.
LSI-97-050 OTB Eg ■ 7/25/97 11:09 PM Page 54 Do not take the Meter apart. Sensitive parts could be damaged and cause inaccurate results. Taking the Meter apart will invalidate the Meter warranty. IMPORTANT: Infection Control Information ■ If you are sharing the ONE TOUCH® BASIC® Meter with another person, always use a new or properly disinfected Test Strip Holder for each person.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 55 Daily Meter Check Holder. Clean the Test Strip Holder and Test Area by following the instructions in the next section. Cleaning the Meter The Meter and Test Strip Holder should be cleaned once a week. In addition, they must be cleaned whenever the Test Area looks dirty or the message Look through the small hole in the Test Strip Holder to make sure there is no lint, dirt, or blood blocking any part of the hole.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 56 Step 1: Remove Test Strip Holder Remove the Test Strip Holder to expose the Test Area. Hold the Meter and place your thumbs on the two raised dots on the Test Strip Holder. Slide the Test Strip Holder toward you.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 57 Step 2: Clean Test Strip Holder Dry completely with a soft cloth or tissue. Step 3: Clean Test Area CAUTION: DO NOT GET WATER INSIDE THE METER. The following cleaning agents will damage the Meter. Wash the Test Strip Holder with soap and water. DO NOT USE: ■ Alcohol ■ Cleansers with ammonia or phenol ■ Windex® or other glass cleaners ■ Abrasive cleansers Clean the underside of the Test Strip Holder and check the small hole for any dirt, blood, or lint.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 58 Check the clear, protective coating over the Meter optics to make sure it is not scratched or damaged. Rub Test Area with a cotton swab or soft cloth dampened with water to remove all blood, dirt, or lint from the Test Area. If necessary, a mild liquid dishwashing detergent mixed with water may also be used. Do not apply full-strength detergent to the Test Area. Be careful not to scratch the Test Area.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 59 Dry the Test Area with a soft, dry cloth or tissue that doesn’t contain lotion or perfume. Remove any lint. R Step 4: Replace Test Strip Holder Press down on the raised dots of the Test Strip Holder until it snaps firmly into place. Press forward on the base, then down on the center of the Test Strip Holder to make sure the small hole over the Test Area is not raised. You are now ready to do a Check Strip test.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 60 The battery should last about one year with typical home use. When the battery is getting low, the display will flash. 6-VOLT Battery Replacement insert insert insert The ONE TOUCH® BASIC® Meter comes with a battery already installed. When the battery must be replaced, use only a Duracell® brand alkaline battery (J-size, 6 volts, part #7K67), which is commonly available in drug and hardware stores.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 61 When Turn the Meter upside down. battry appears on the display, the Meter will no longer give results and you must replace the battery before you can perform another test. A dead battery or battery removal will not affect the previous test result stored in the Meter’s memory, as long as the battery is replaced within 30 days. With your thumb, firmly press in and up on the ridges on the side of the battery door until the door releases from the clasp.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 62 6-VOL T Remove the old battery and discard properly. Insert a new J-size battery in the battery compartment. To replace the battery door, insert the tabbed side of the door into the slot in the battery compartment.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 63 With your thumb, press in and down on the ridged side of the battery door until it snaps shut.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 64 9. QUESTIONS AND ANSWERS under control. Your ONE TOUCH BASIC System helps you check whether or not you are in control by monitoring your blood glucose levels. In the following pages, we’ve provided answers to some of the most common questions people have about home blood glucose monitoring and the ONE TOUCH® BASIC® Meter. If you have a question that isn’t covered here, consult your Health Care Professional or call LifeScan Technical Services.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 65 Q. How often should I test my blood glucose? variations provide useful clues about your condition and how well your treatment program is working. A. This is best decided by you and your Health Care Professional. It may vary according to your age, the type of diabetes you have, any medication you are using, whether you are ill, and any physical or emotional changes in your life.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 66 Keep your Meter clean at all times. Make sure the Meter code number is set to match the code number on the package of Test Strips you are using. Use enough blood. The entire Test Spot must be completely covered with blood. Touch the drop of blood lightly to the middle of the Test Spot to fill the entire Spot. The drop of blood should form a round, shiny circle over the entire Spot, and stay wet during the entire test.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 67 your Health Care Professional. If your blood glucose test results remain consistently high, contact your Health Care Professional. Q. What do I do if test results still seem too low or too high? A. Remember, the test results you obtain will vary from time to time depending on food intake, insulin dose, exercise, etc. Ask yourself the following questions: Q. What are the expected blood glucose values associated with well-controlled diabetes? A. Fasting: 3.3–7.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 68 In addition to the common factors listed above which cause real variations in blood glucose results, there are certain abnormal conditions which can interfere with the accuracy of your blood glucose monitoring results. These limitations are described below and should be discussed with your physician: Severe dehydration and excessive water loss may cause inaccurately low results.
LSI-97-050 OTB Eg ■ ■ ■ ■ 7/25/97 11:09 PM Page 69 Vomiting and diarrhea Prescription drugs, e.g., diuretics Inability to recognize or respond to “thirst” sensations Sustained uncontrolled diabetes in this Owner’s Booklet as the cause, you should contact a physician immediately! Never make significant changes to your medication program or ignore physical symptoms without consulting a physician. Whenever inadequate fluid intake or excessive water loss occurs, CONSULT A PHYSICIAN IMMEDIATELY.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 70 Q. Why don’t my Meter results match the results that I got from my doctor’s laboratory exactly? 10%–12% lower than serum or plasma values from the same blood sample (at average hemocrit levels of 41%–45%). To adjust for this difference, divide the laboratory value by 112% or 1.12.7 A.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 71 There is a difference in the glucose levels of venous blood (usually drawn from a vein in your arm) and capillary blood (blood from a fingerstick). After a meal, the glucose levels in the capillaries may be 1.1–3.9 mmol/L higher than the levels in venous blood.6 If your lab test is taken while you are in the fasting state (before breakfast), this difference will be minimal (0.1–0.3 mmol/L).
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 72 Dimensions: Length—12.07 cm (43/40) Width—6.70 cm (25/80) Height—2.70 cm (11/160) Weight—approx. 135 g (4.8 oz.) [including battery] 10. SPECIFICATIONS Performance Characteristics See the ONE TOUCH® Test Strip package insert for ONE TOUCH® System accuracy and precision. Operating Temperature Range: 15°C–35°C (59°F–95°F) Power Supply: One Duracell® alkaline battery, size J (6V) part # 7K67 is included with Meter.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 73 Guarantee and Warranty replace it with a new or rebuilt Meter or equivalent product free of charge. 30-Day Money-Back Guarantee If you are not fully satisfied with the ONE TOUCH® BASIC® System, a full refund may be obtained by calling a LifeScan Technical Representative at the toll-free number within 30 days of purchase. You must return the ONE TOUCH® BASIC® Meter and a copy of your receipt to receive a refund.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 74 The warranty policy does not apply to the performance of the ONE TOUCH® BASIC® Meter when used with any test strip other than ONE TOUCH® Test Strips, or when the ONE TOUCH BASIC Meter or Test Strips are changed or modified in any way. This warranty is in lieu of all other warranties, express or implied, including any implied warranty of merchantability or fitness for a particular purpose.
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 75 11. INDEX Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Blood Glucose Values (expected) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Blood Sample Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Check Strip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 76 Coding the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Control Solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11, 17 Displaying Last Test Result . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 38 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43–52 Hematocrit . . . . . . . .
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 77 Precautions and Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Questions and Answers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 Testing Your Blood . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27–34 Test Procedure Messages . . . .
LSI-97-050 OTB Fr 7/25/97 11:00 PM Page 86 Toll-Free customer service: Appels sans frais: Canada 1 800 663-5521 U.S.A. 1 800 227-8862 Burnaby, B.C. V5C 6C6 *05160052A* AW *05160052A* Printed in U.S.A. Imprimé aux États-Unis © 1994 LIFESCAN INC.