Product Info
Page 260 of 264
6. CHL50 (MEDLINE
®
CARELINE
®
)
The CHL50 charger is an important part of the LIFT50 family. It functions as a wall
charger when mounted on the wall, but also as a more mobile charger simply placed
horizontally on its back on any surface.
Usage:
• Usage temperature: +5 °C to 40 °C
• Storage temperature: -10 °C to 50 °C
• Relative humidity: 20% to 80% - non-condensing
• Atmospheric pressure: 700 to 1060 hPa
• Height above sea level: Max. 3000 meters
• Nominal current draw: Max. 350 mA
• Power consumption (standby): < 0.5 W
• Power consumption (charging): Max. 25 W
• Approvals: (pending) IEC 60601-1
ANSI/AAMI ES60601-1
CAN/CSA-22.2 No. 60601-1
IEC 60601-1-2
PSE
LED 1 - LED 2 - LED 3
LED state
(Not listed = off)
States while charging
LED 1 - 3 constantly on
With battery or if no battery
mounted on charger: 90 - 100%
LED 1 + 2 constantly on
LED 3 flashes slowly
65 - 90%
LED 1 constantly on
LED 2 flashes slowly
40 - 65%
LED 1 flashes slowly 0 - 40%
LED 1 + 2 + 3 flash slowly
Charging stopped due to low
battery temperature, high bat-
tery temperature or other error
conditions
No light in LEDs
Charging stopped due to lost
communication to battery
LED indicators