Users Manual

Page 173 of 264
CBJ1 CH1 CH1 Leg spreader
CH2
Up Down Up Down
Min. Max. Max. Min. Min. Max. Max. Min. Min. Max.
CBJ1001N 6.3 8.3 5 3.3 * * * * - -
CBJ1001H 9 11.5 7 4.7 6.3 8.8 7 4.7 - -
CBJ1002N 6.3 8.3 5 3.3 * * * * 2 2.9
CBJ1002H 9 11.5 7 4.7 6.3 8.8 7 4.7 2 2.9
CBJ1004H 5.25 6.3 3.85 2.75 5.2 7.1 5 3.3 4.7 6.2
CBJ1005N 6.3 8.3 5 3.3 * * * * 4.7 6.2
CBJ1005H 9 11.5 7 4.7 6.3 8.8 7 4.7 4.7 6.2
*Current cut-off depends on motor type due to pulse width modulation.
Current cut-off for CBJ1
CBJ2 CH1 Leg spreader CH2
Min. Max. Min. Max.
CBJ2001N 7 9 - -
CBJ2001H 9.8 11.8 - -
CBJ2002N 7 9 2 2.9
CBJ2002H 9.8 11.8 2 2.9
CBJ2003H 4.9 5.9 - -
CBJ2004H 4.9 5.9 4.7 6.2
CBJ2005N 7 9 4.7 6.2
CBJ2005H 9.8 11.8 4.7 6.2
CBJ2006H 9.8 11.8 - -
CBJ2007H 9.8 11.8 2 2.9
CBJ2008H 9.8 11.8 4.7 6.2
Current cut-off for CBJ2
The minimum and maximum values stated above refer to the tolerance range, not the adjustment range.
The current cut-off tolerance is +/- 0.2 A depending on the ambient temperature (20 �C)
To avoid injury, all control boxes with variable current cut-off are preset to < 4 A, unless otherwise specified.
When plugging a mono jack plug into channel 1 the current cut-off will be as the high speed table.