Users Manual

Page 175 of 264
The control box CBJ Care is part of the JUMBO system. JUMBO is a
modular system combining an actuator, control box (CBJ Care), battery
(BAJ1, BAJL), wall charger (CHJ2), control box prepared for external
charger by use of wall-plug charger CH01, and a hand control in a flexible
solution, specially developed for patient lifts.
The complete system contains a series of features which meet the patient’s
need for a safe and comfortable lift.
CBJ Care is available in 3 versions, one with LEDs, one with a display and
a third without display and LEDs.
It is possible to have control buttons on the front cover to have an easy
control option if the hand control is missing. Furthermore it is possible to
have 3 channels via a T-cable in channel 1. The 3
rd
channel for tilt function
adds value for the patient and the caregiver.
Usage:
CBJ Care with internal charger: Nominal current draw max. 400 mA
Power consumption (standby) max. 2.5 W
Power consumption (charging) max. 19 W
Duty cycle: Max. 10 % or 2 min.continuous use then 18 min. without use
Ambient temperature: +5° to +40°
Storage temperature: -10° C to +50° C
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa
Approvals: IEC60601-1, IEC60601-1-6
ANSI / AAMI ES60601-1,
CAN/CSA-22.2 No 60601-1
4. CBJ-Care (MEDLINE
®
CARELINE
®
)
Instructions for uses
Before start-up we recommend to reset the service counters – days and cycles until next service visit. To reset press the up and down button on the
control box or the hand control for 5 seconds. An audio signal will confirm the resetting.
When charging, the CBJ Care will not be able to operate any actuators.
It is not possible to use other battery types than BAJ1 or BAJL with the CBJ Care.
Use only original LINAK mains cables to ensure proper connection to internal charger.
The green battery indicator (100% to 50% capacity remaining) will light up during charging even though the battery is not fully charged.
It is necessary to use the CHARGE” LED to indicate whether or not the battery is fully charged (when using internal charger).
The CHARGE indicator will light up during charging and turn off when the battery is fully charged.
When resetting the CBJ Care or updating other settings than using learn mode, the CBJ Care must not be disconnected from the battery and the
emergency stop must not be activated within a time span of at least 10 seconds. This is to ensure the correct storage of the new values to the
memory banks of the CBJ Care.
Hot Plugging:
Removing or adding any OpenBus™ cables is not allowed when the control box is on power via mains supply or battery!
If necessary anyway, follow the below procedure:
1. Remove mains or battery and wait 5 sec.
2. Mount or dismount the required cables
If this procedure is NOT followed it may result in a damaged OpenBus™ driver circuit.
The risk of a damaged circuit increases if the accessory has a high start current (in rush current).
Emergency lowering/lifting:
By use of BAJ1, the lifting arm can be lowered by pressing e.g. a pen in the hole or use the control buttons, if present.
This is a permitted method of lowering/lifting.
The emergency lowering/lifting “buttons” work as normal hand control buttons (you do not get extended functionality by
using these when the battery is low).
By use of BAJL, please be aware that loss of power might happen due to the battery deep discharge protection.
This will only happen by continuous use of the battery despite warning.
Recommendations
Warnings
In order to avoid injury, the emergency-stop should be activated in (all) shipping situations.