Total cycles done on channel 1
Total work done on channel 1
Total number of overloads (channel 1)
Days since last service/Days between services
If “No days are chosen for service interval then the display will show Days since last service /-.
Overload Channel 1 only:
When overload occurs (according to the pre- defined current cut off limit) the overload symbol will be shown on the display.
The ‘MAX’ text will blink 3 times and the overload symbol will be shown for 10 seconds in total.
Service information read-out
Basic service information can be read out on the display. To get the service information on the display please press the lifting arm up button (only ½
second press). The information will be shown for ½ minute or until other buttons are activated.
Special care should be taken when mounting the CBJ Care.
As long as the the CBJ Care is mounted correctly then the CBJ Care complies to IPX4. If the CBJ Care is mounted incorrectly then water will
gather around the screw holes resulting in non-compliance with IPX4!
If the control box is equipped with emergency stop, the stop button must be activated in cleaning situations in order to comply with IPX4.
The battery pack BAJ1 or BAJL must NOT be removed in cleaning situations, doing so could result in non-compliance with IPX4.
If the CBJ Care is fitted with external charger option (DC power connector), the protection plug ex. 00918174 must always be inserted to
ensure IP protection, if the port is not used.
IP rating only applies when the battery is connected to the control box.
Mounting of CBJ Care
Recommendations for use of learn mode function:
The purpose of using learn mode function is to adjust the lift to no more than 1.5 times the max. load. The actuator will not stop exactly at the load
it has been adjusted to as the actuator uses less current when its components have been run in. When the max. current value has been registered
using the learn mode function, the control box will be able to use max. current +10 %. This ensures that the lift is capable of lifting the set load,
however it cannot lift more than 1.5 times of the set load.
When registering current limits, be aware to use a defined set of actuator and control box
The ambient temperature must be approx. 20 ºC
The difference between the highest and lowest load must not be more than max. 10 %
To activate the learn mode function, use the special hand control (HB7x235-00)
If an actuator or CBJ Care is exchanged, it is necessary to reset the max. load to ensure the correct cut-off value for the new system
Always use fully charged batteries for learn mode procedures
A max. cut-off value of 11 Amp can be registered (stored)
The tolerance for preset current cut-off is: +/- 1 Amp
The current cut-off value can be reset by means of the learn mode function, however this is not in accordance with EN10535