8. Information on specific accessories
If the actuator is to be equipped with accessories, these must be specified when ordering the actuator from LINAK. There are the following possibilities:
1) TR6/TR7 External transformer
If the TR6 or TR7 fixed cable connection becomes damaged the transformer must be replaced.
1. BA16 Lead acid (MEDLINE
®
CARELINE
®
)
The battery box BA16 is developed for use together with the LINAK CA and CO control box
series to support power backup.
Usage:
Compatibility: Battery back-up for COxx and CAxx
Duty cycle: 10%, 2 minutes continuous use followed by 18 minutes not in use
Charging: Via integrated charger
Charging time: Approx. 6 hours
Recharging during storage:
Battery recharging no later than 6 months after production date stated on the label
Operating temperature: +5 °C to +40 °C
Storage temperature: -10 °C to +50 °C
The batteries must be stored in an applicable storage room to avoid direct sunlight
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa
Height above sea level: Max. 3000 meters
Service: Battery replacement
Approvals (pending): IEC60601-1, ANSI/AAMI ES60601-1, CAN/CSA-22.2 No. 60601-1
UL tested in accordance with UL60601-1 (pending)
LED functionality:
LED Indication of operation
Solid yellow Charging (battery not ready)
No LED light Fully charged (battery ready)
Flashing yellow Error during charging
Buzzer functionality:
The buzzer will make a warning when a button on the hand control is pressed and the battery capacity is low. The buzzer can also be activated by an
intelligent control box to signal other conditions. This must be specified in the control box software.