Users Manual

Page 183 of 264
Recommendations:
Do not exceed the storage temperature as it will shorten the product life and reduce performance.
Allow the battery to settle to room temperature before use.
Do not exceed the duty cycle 2/18 as it will shorten the life, reduce performance, and eventually activate
overcurrent protection.
BA16 is not intended for use in outdoor applications.
If the battery is completely discharged, then recharge the battery before storage.
Inspect at regular intervals that the ventilation aperture is positioned correctly and is intact throughout its length.
Safety feature
BA16 contains overcurrent protection for safety and to protect itself from being damaged due to excessive use.
When current protection is activated, no power output will be available.
Warnings
Loss of power might happen due to activation of overcurrent protection. In this event, there may be no warning and the application may not be able
to move when expected.
Defective or damaged batteries may leak acid and adequate precautions must be taken during handling and transportation.
Do not open the battery case as damage to the cell or circuitry may develop excessive heat.
It is important for users to read the guidelines in the User Manual Linear Actuators and Electronics”.
Do not short circuit the battery.
Use the specified internal charger only.
If disposed to fire, the battery may explode.
The battery box BA16 itself may not be combined with an external charger.
If product caution is not clearly visible on the final application at low light intensity, the above mentioned warnings must be integrated in the application
manufacturer manual.
The application manufacturer must test the application and ensure that neither intended nor unintended use exceeds the battery specification. The
application manufacturer must assure other means of movement, e.g. quick release or manual lowering in case of battery failure.
Compatibility:
The BA16 has a built-in charger and is therefore not able to operate with control boxes with charger. Be aware that the BA16 is only
compatible with CAxx and COxx.
BA16 safety:
LINAK batteries for medical use are designed and manufactured to be safe throughout the product life. LINAK has performed various battery tests in
normal use, abuse, and failure situations to verify design and production methods. These tests have not shown any unacceptable risks.
The batteries are UL-tested to verify the safety of the design and to obtain a safety certificate from an independent organisation. This means that UL
regularly inspects the factory to check that standards are complied with.
Mounting instructions:
BA16 must be mounted with M4 screws due to the battery weight.
Make sure the surface touching the BA16 mounting surface is flat and use all 4 screws.
The diameter of the screw cap must be maximum 8mm.
d
1
M4
d
2 max
8
k 4
According to ISO 2009
Cable Mini-fit (4 pole) with angle to Mini-fit (4 pole) straight
For cable details see chapter 4.1.4/see cable configurator
Cable lock 0273044
Cable
Created By Lian Wang 2019-01-09
Modified By Lian Wang 2019-01-09
SCALE : 2:1
Drawing No.: 0273044-SALES
Created By Lian Wang 2019-01-09
Modified By Lian Wang 2019-01-09
SCALE : 2:1