Users Manual
Page 184 of 264
2. BA18 (MEDLINE
®
CARELINE
®
)
The BA18 is a cost-effective lead acid battery without integrated charger that can be used in combination with
CO53 and with the long-established control box range, of which many are now legacy products.
Usage:
• Compatibility: CO53, CB6P2
Legacy products : CB6,CB7, CB9, CB12, CB14, CB18, CB6S
NOTE: (only specific versions may be compatible)
• Ambient temperature: +5°to +40°C
• Charging: Via LINAK control box with integrated charging circuit
• Charging time: Approx. 6 hours depending on built-in control box charger
• Recharging during storage: Battery recharging no later than 6 months after production date stated on the label
• Operating temperature: +5 °C to +40 °C
• Storage temperature: -10 °C to +50 °C
• Relative humidity: 20% to 80% non-condensing
• Atmospheric pressure: 700 to 1060 hPa (3000 m)
• Meters above sea level: Max. 3000 meters
• Approvals: IEC60601-1, ANSI/AAMI ES60601-1, CAN/CSA-22.2 No. 60601-1
Warnings:
• The battery case is only to be opened by authorised staff as incorrect handling may compromise the IP protection.
• Take care to always keep the venting channels free. Mounting plates must be rigid to prevent blocking of the venting channels.
• Do not use third party chargers.
Recommendations:
• Allow the battery to settle to room temperature before use.
• The batteries must be stored in an applicable storage room to avoid direct sunlight.
To ensure free passage of gasses when the battery is mounted on a flat surface the back side of the battery has been supplied with venting channels
see below figure.
Venting channels and membrane on BA18:
Check with regular intervals that the venting channels are unblocked.