Page 184 of 264
2. BA18 (MEDLINE
®
CARELINE
®
)
The BA18 is a cost-effective lead acid battery without integrated charger that can be used in combination with
CO53 and with the long-established control box range, of which many are now legacy products.
Usage:
Compatibility: CO53, CB6P2
Legacy products : CB6,CB7, CB9, CB12, CB14, CB18, CB6S
NOTE: (only specific versions may be compatible)
Ambient temperature: +5°to +40°C
Charging: Via LINAK control box with integrated charging circuit
Charging time: Approx. 6 hours depending on built-in control box charger
Recharging during storage: Battery recharging no later than 6 months after production date stated on the label
Operating temperature: +5 °C to +40 °C
Storage temperature: -10 °C to +50 °C
Relative humidity: 20% to 80% non-condensing
Atmospheric pressure: 700 to 1060 hPa (3000 m)
Meters above sea level: Max. 3000 meters
Approvals: IEC60601-1, ANSI/AAMI ES60601-1, CAN/CSA-22.2 No. 60601-1
Warnings:
The battery case is only to be opened by authorised staff as incorrect handling may compromise the IP protection.
Take care to always keep the venting channels free. Mounting plates must be rigid to prevent blocking of the venting channels.
Do not use third party chargers.
Recommendations:
Allow the battery to settle to room temperature before use.
The batteries must be stored in an applicable storage room to avoid direct sunlight.
To ensure free passage of gasses when the battery is mounted on a flat surface the back side of the battery has been supplied with venting channels
see below figure.
Venting channels and membrane on BA18:
Check with regular intervals that the venting channels are unblocked.