3. BA19 Lead acid (MEDLINE
®
CARELINE
®
)
The BA19 lead acid backup battery has been developed specifically for
use with the new control boxes CA30/CA40 and CO61. It is a compact
and cost-efficient battery with built-in charger and cable management.
Usage:
Compatibility: Battery backup for CA/CO control box platform
Duty cycle: 10%, 2 minutes continuous use followed by 18 minutes not in use
Charging: Via integrated charger
Charging time: Approx. 6 hours
Recharging during storage: Battery recharging no later than 6 months after production date stated on the label
Operating temperature: + 5 °C to + 40 °C
Storage temperature: - 5 °C to + 40 °C
The batteries must be stored in an applicable storage room to avoid direct sunlight
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa (3000 m)
Height above sea level: Max. 3000 meters
Service: Battery cells cannot be replaced as the battery cover cannot be closed properly afterwards
Approvals (pending): IEC60601-1:2005 3
rd
edition,
ANSI/AAMI ES60601-1: 2005, 3
rd
edition
CAN/CSA-22.2 No. 60601-1:2008
LED functionality:
What does the LED indicate?
Buzzer functionality:
The buzzer will make a warning when a button on the hand control is pressed and the battery capacity is low.
The buzzer can also be activated by the control box to signal other conditions. This must be specified in the control box software.
LED Indication of operation
Solid orange Charging (battery not ready)
No LED light Fully charged (battery ready)
Flashing yellow Error during charging
Mounting bracket (frame flat) - article no. 1015W1001: Mounting instructions:
BA19 must be mounted with attachment
screw and mounting bracket (see below)
due to the battery weight.
Screw M4 countersunk torque
1.1 Nm +/- 0.1 Nm.
d M4
d
k
8.4
k 2.7
According to ISO 2009
d
k
d
k
45°