Recommendations:
• The MJB must be mounted on an even surface
• The locking surface must be free of other material
• Always use locking mechanism and O-ring
• Sockets not used must be fitted with blind plugs to ensure the IP degree
• HOT PLUGGING
Removing or adding any OpenBus cables are not allowed when the control box is powered by mains supply!
If needed anyway follow the below procedure:
1. Remove mains and wait 5 sec.
2. Mount or dismount the required cables
If this procedure is NOT followed it may result in a damaged OpenBus driver circuit.
The risk of a damaged circuit increases if the accessory has a high start current (in rush current).
• When using USB cable (0834000) or modular plug cable (0964399) with open end, it is up to the customer to maintain the IP degree.
• Do not use 2 MJB5 Plus variants with same device ID on the OpenBus™.
This will cause conflicts and the SDT is not able to identify the different products attached.
• Before the final functional test in the production, is it important that the system is re-powered. This is to make sure that all items have been
detected on the OpenBus.
• We recommend that the end user makes a regular test procedure, in order to prevent failures and hazardous situations on the system, e.g.
squeezed cables. The MJB5 Plus is not able to detect defective 3
rd
party products.
• LINAK only takes responsibility for LINAK products, not 3
rd
party products.
Please pay attention to the “Patient Environment” Clause 3.79 - IEC60601-1.
• There can be a risk of conflict with other OpenBus accessories, such as HB, etc. it is therefore recommended to make a system/bit overview.
• When connecting 3
rd
party products to LINAK systems, the customer must take necessary precautions against Electrostatic Discharge (ESD).
Exposure to harmful ESD must be avoided.
• 3
rd
party products must be designed with the following isolation:
Minimum 1 MOPP (creepage distance/clearance according to IEC 60601-1).