Users Manual

Page 167 of 264
19. HD80 JUMBO (MEDLINE
®
CARELINE
®
)
The HD80 JUMBO is a hand control with an optimised ergonomic design and functions that
are activated via dome buttons.
Usage:
Usage temperature: 5º C to 40º C
Storage temperature: -10º C to +50º C
Compatibility: Only compatible with CBJ Care
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa (3000 m)
Height above sea level: Max. 3000 meters
Flammability rating: UL94-V2
Approvals: IEC60601-1, ANSI/AAMI ES60601-1 and
CAN/CSA-22.2 No 60601-1
Warnings
Do not sit or lie on the hand control. It can cause unintended movement of the application.
Always use O-ring on connectors and cable locks.
There is a risk that items with internal magnet for mounting instead of hook can disturb cardiac pacemaker functions, implantable cardioverters,
defibrillators or magnetic implants.
Recommendations
Clean the hand control regularly to ensure good hygiene standards.
When a defective HD80 is replaced, check that the new HD80 has exactly the same specification and functionality.
Do not submerge the hand control under water.
Unless otherwise specified or agreed by LINAK, the hand control is only intended to be used on LINAK systems.
When changing hand controls for OpenBus™, the power must be switched off.
It is recommended to check the hand control and cable for damage and holes made by violent handling before washing the bed or at least once a year.
In order to maintain the flexibility of the cables, it is important that a coiled cable is placed in such a way that the cable’s own weight does not strain
the coil during the washing process.
Hand controls with magnets:
If hand controls with magnet are hooked on a smooth surface, a movement or twisting of the cable, for instance during transport, can cause the hand
control to move and result in damage, if the cable is squeezed somewhere.
The force of the magnet depends on the lacquering thickness, the lacquering type, stickers, steel thickness etc. The customer has the responsibility
to verify that the holding force on the application is acceptable.
The user/operator is responsible for evaluating any potential risk caused by the use of magnets.
It is recommended to have a parking spot for the hand control on the application where the customer ensures that the hand control does not fall off.
The HL70 is a hand control with integrated locking function, where a selec-
tive locking of the different functions is available by use of a special key.
The HL70 is an alternative to the HB70 combined with an attendant Control
Panel (ACM, ACL, etc.)
Usage:
Exchangeable with HB70
Operation temperature: +5 °C to +40 °C
Storage temperature: -10 °C to +50 °C
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: Max. 3000 meters
Compatible with many LINAK control boxes
Approvals: IEC 60601-1
IEC 60601-1-6
ANSI/AAMI ES60601-1
CSA CAN/CSA-C22.2 NO. 60601-1
20. HL70 (MEDLINE
®
CARELINE
®
)
Recommendations
To switch between locked and unlocked position a small knob between the two push buttons has to be turned 20° by use of a special key.
The key is for the use of the nursing staff only, there are two types, one is made of plastic the other metal.
For all types: Attention should be given to ensure that the channels shown correspond to the channels available on the chosen control box.
The HL70 must hang vertically from its hook during the washing process. In order to maintain the flexibility of the cables, it is important that
a coiled cable is placed in such a way that the cable weight does not strain the coil during the washing process.