Users Manual

Page 169 of 264
The BAJL Li-Ion battery pack has been specially developed for use with the JUMBO system for patient lifts and sit to stand lifts. It is a low-weight battery
with reliable and high performance.
Usage:
Compatibility: CBJ Care, COBO, CHJ2 and CH01
Duty cycle:
BAJL003xxxxxxxx: 10 % (2/18 min.) at max. current draw 10 Amp (ambient temperature ≤ 30 ºC)
10 % (2/18 min.) at max. current draw 8 Amp or
5% (1/19 min.) at max. current draw 10 Amp (ambient temperature > 30 ºC)
BAJL004xxxxxxxx: 10 % (2/18 min.) at max. current draw 10 Amp (standard ambient temperature recommendations)
Charging: Via JUMBO wall charger CHJ2 or via JUMBO control box with integrated charger
Charging state: Maximum 30% when shipped from LINAK
Recharging during storage: Recharge the battery 6 months at the latest after production date stated on the label
Operating temperature: +5 °C to +40 °C
Charging temperature: +10 °C to +40 °C
Charging time: Type 3: 3 to 4 hours - Type 4: 6 to 8 hours
Storage temperature: -10 °C to +40 °C (+10 °C to +25 °C - recommended)
The batteries must be stored in an applicable storage room without direct sunlight.
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa (3000 m)
Height above sea level: Max. 3000 meters
Approvals: IEC60601-1:2005 3
rd
edition,
ANSI / AAMI ES60601-1:2005, 3
rd
edition,
CAN / CSA-22.2 No 60601-1:2008,
IEC62133 2
nd
edition,
UL2054, 2
nd
edition
PSE (pending)
UN38.8, 6
th
edition (needed for transport of lithium batteries)
7. Information on specific JUMBO
TM
2. BAJL Li-Ion (MEDLINE
®
CARELINE
®
)
1. BAJ (MEDLINE
®
CARELINE
®
)
The battery pack BAJ has been specially developed for use with the JUMBO
system.
They are easy to exchange through an integrated snap system, and can
easily be mounted on the mounting brackets.
Usage:
BAJ is a part of the JUMBO system. It is compatible with CBJ1/CBJ2, CHJ2, CBJ Care and COBO
Duty cycle: 10 % or 2 min. continuous use then 18 min. not in use
Ambient temperatures: +5 °C to +40 °C
Storage temperature: -15 °C to +40 °C
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa
Height above sea level: Max. 3000 meters
Flammability rating: UL94-V0
Approvals: IEC60601-1, ANSI/AAMI ES60601, CAN/CSA-22.2 No 60601-1
Charging:
BAJ can be charged by
Charger CHJ2
Control box CBJ1/2, CBJ Care, COBO
BAJ with integrated DC plug can also be charged by use of the external charger CH01
Warning
Check at regular intervals that the ventilation hole is undamaged and intact. The construction of the ventilation stub permits battery
gasses to get out, but it does not permit penetration of water.