Installation Instructions

Page 188 of 194
13. SLS (MEDLINE
®
CARELINE
®
TECHLINE
®
)
LINAK has designed a switch that can be mounted in connection with the 24V DC actuators: LA12, LA22, LA28,
LA28S, LA31, LA32, or LA34 and a control box on e.g. a bed frame.
The Safety Limit Switch (SLS). The SLS Switch is primarily used as a Limit Switch in systems consisting of a control
box, LA28/28S and/or LA32.
As an example the SLS can be installed on the actuator cable where it disconnects the actuator in either inward or
outward direction when activated. The SLS can also be used as a signal control directly connected to a control box.
The Under Bed Light (UBL) can be mounted under the bed to provide a discrete guiding
light when the patient leaves the bed during night. With an Under Bed Light it is easy to
switch the light on and off by means of a handset or controlling it automatically with the
Out of Bed functionality of the MJB8.
14. Under Bed Light (MEDLINE
®
CARELINE
®
)
Usage:
Operating temperature: + 5 °C to + 40 °C
Storage temperature: - 10 °C to + 50 °C
Atmospheric pressure: 700 to 1060hPa (3000 m)
Approvals: IEC 60601-1, ANSI/AAMI ES60601-1,
CAN/CSA-22.2 No 60601-1
Connectivity: To be connected to MJB5 and MJB8
Recommendations
The Under Bed Light (Item P/N.: 0964135) must be mounted on the bed with metal screws in order to maintain ESD protection.
The Under Bed Light (UBL2) provides a powerful light with
a good distribution. The UBL2 is to be used for beds within
hospitals, nursing homes and in homecare.
The Under Bed Light makes it easier for patients and other
people in need of care to find their way at night in the dark
to prevent falling accidents and to make them feel safe.
15. Under Bed Light 2 (MEDLINE
®
CARELINE
®
)
Usage:
Operating temperature: + 5 °C to + 40 °C
Storage temperature: - 10 °C to + 50 °C
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa (3000 m)
Height above sea level: Max. 3000 meters
Approvals: EN62471, IEC60601-1
Recommendations and Precautions
If 2 or more UBL2 products with dimming function are connected to the same application, it is recommended to have a factory reset key to be
able to synchronize the dimming direction and light intensity if one of the UBL2s is replaced.
Screw holes in application are needed for mounting. Inform the customer to use M4 Ø12 screws with Ø12 washer, when mounting the UBL2.
Max. torque 2.5 Nm.
• Hot-plugging:
Removing or adding any OpenBus™ cables is not allowed when the CB is powered by mains supply!
If needed anyway, follow the below procedure:
1. Remove mains and wait 5 sec.
2. Mount or dismount the required cables
If this procedure is NOT followed, it may result in a damaged OpenBus™ driver circuit. The risk of a damaged circuit increases if the accessory
has a high start current (in rush current).
There can be a risk of conflict with other OpenBus™ accessories, like HB, ACP, etc. when using the OpenBus™ UBL2, it is therefore
recommended to make a system/bit overview.
Always use locking mechanism and O-ring
Sockets not used must be fitted with blind plugs to ensure IP degree
The UBL2 must be mounted on a plane surface and casing must not be subject to impact or any kind of stress.