Installation Instructions
Page 189 of 194
16. WET Sheet (MEDLINE
®
CARELINE
®
)
The WET Sheet is an OpenBus™ product and is a part of the WET Detection solution.
The WET Detection solution consists of the WET Sheet, the MJB8 and a WET Sheet cable.
Recommendations
• The WET Sheet is a wearing part which is not covered by the standard LINAK warranty
The WET Sheet warranty is valid until the product has been put into use
• The WET Sheet is not a stand-alone article and must be included as part of an application e.g. the MJB8
• The WET Detection will not detect liquid without conductivity e.g. demineralised water
• The WET Sheet is not intended for outdoor use
• The WET Sheet is not a life-supporting unit
• Mechanical pressure on the WET Sheet can destroy the thread and connectors
• It is recommended to change the WET Sheet cable after 500 connections
• The sheet and connector must be placed properly to minimise the risk of bedsore
• The WET Sheet connector is not to be exposed to direct moisture etc. urine or sweat.
Place the connector away from the user e.g. on the side of the bed
• To ensure the right WET Detection functionality do not use layers on top of the WET Sheet, e.g. a turning sheet
• The recommended temperature for washing is 60 °C. Washing at 85 °C is acceptable, however,
this will reduce the lifetime
• Tumble drying at medium heat, however, line drying is recommended
• The typical number of washing cycles is up to 50, but will depend on the washing conditions
• The WET Sheet should be washed before use
17. QLCI2 (MEDLINE
®
CARELINE
®
)
The Quad Load Cell Interface 2 (QLCI2) for the
hospital and care segment is a scale system
accessory with weighing capabilities and Out of
Bed functionality.
The housing, makes the QLCI2 easy to mount
by unique slide-on brackets and has an IPX6
Washable DURA
TM
ingress protection.
The QLCI2 supports the LINAK OpenBus
TM
system offering a high level of customisation.
LINAK A/S delivers the OpenBus™ system only
and is not responsible for any products (i.e.
products from 3
rd
party suppliers) other than
LINAK products or the compatibility of such
products with the LINAK OpenBus™ system.
Usage
• Operation temperature: + 5 °C to + 40 °C
• Storage temperature: - 10 °C to + 50 °C
• Relative humidity: 20% to 80% - non-condensing
• Atmospheric pressure: 700 to 1060 hPa (3000 m)
• Height above sea level: Max. 3000 meters
• Approvals IEC60601-1
ANSI/AAMI ES60601-1
CAN/CSA-22.2 No 60601-1
EN 45501 / OIML R76
EU type examination according to 2014/31/U
• Compatibility: All OpenBus
control boxes
• Flammability rating: UL V0
• Latex free: Yes