Product Drawings

Page 151 of 194
Recommendations:
The application manufacturer must ensure a proper installation of the FPP in the application which is convenient for the end user.
To ensure proper activation, the lock above the housing must be properly locked by turning it clockwise.
The application manufacturer must use the correct torque for the slotted set screw of the bracket to ensure a stable positioning of the FPP.
The application manufacturer must consider the bracket position carefully. If the FPP is mounted on a moveable part, it will move and might
touch the patient or parts of the application. If, however, mounted on a fixed part, the FPP might not be within the reach of the patient.
The end user must not apply a torque to the FPP housing of more than 8 Nm between the flexible arm and the panel.
The end user must not bend the FPP arm to a radius smaller than 105 mm.
The FPP must never be used as a handle for moving the application.
The end user must be informed that the FPP must not be used for other purposes (such as table, handle) than intended.
The end user must take care that the FPP does not touch items or persons when the application is moved.
Warnings:
The FPP must be placed readily accessible for the patient. Never let the FPP hand out of the bed.
Never use the FPP as a handle.
Do not use sharp devices to activate buttons on the FPP.
Never use the FPP as support device. The FPP must not be used as table or notepad, nor can it be used to hang objects on.
As illustrated in the pictures below the panel itself can be moved and angled in a number of positions. The arm can also be bent to move it closer or
move it further away from the user.
The lock function
Between the arm and the panel there is a lock/unlock function, (a hose type connection).
It enables the user to turn the panel into a preferred position.
Locking of the panel:
Turn the panel to a preferred position. With one hand on the panel turn the hose clockwise with the other hand.
The panel is fully locked when it cannot be turned.
Unlocking of the panel:
With one hand on the panel, turn the hose counterclockwise with the other hand until the panel can be moved freely.