The EBC - Electronic Brake Control - is designed for use together with 3
rd
party castors
and the control systems CB OpenBus
TM
, including JUMBO Care. The EBC can replace
traditional (mechanical) central locking systems for castors on healthcare beds or medical
applications. This new system offers an increased freedom of application design as there is
no need for a mechanical connection between the castors.
Usage:
Operating temperature: 5 °C to 40 °C
Storage temperature: - 10 °C to 50 °C
Compatibility: CBJC, CB6 OBF, CB6 OBMe, CB16 or CB20
Fallshaw EBC system approved according to IEC60601-1:2005 3
rd
edition approved,
ANSI / AAMI ES60601-1:2005, 3
rd
edition, and CAN / CSA-22.2 No 60601-1:2008
Manner / TENTE approvals are pending
8. EBC - Electronic Brake Control System (MEDLINE
®
CARELINE
®
)
Recommendations
The customer shall ensure proper connection between the EBC and the castors. Missing or interrupting some of the individual wires between the
EBC and the castor may under some circumstances cause damage to the internal circuitry of the EBC.
Always use locking mechanism and O-ring on cables.
If any open sockets, they must be fitted with blind plugs to ensure IP degree.
Removing or adding any OpenBus™ cables is not allowed when the control box is powered (hot plugging).
Before installation/service
Stop the application.
Remove battery power cable and OpenBus™ connection then possible castor connection.
• Service system.
After installation/service
Reconnect castor connection, then OpenBus™ connection, battery power cable, then control box mains.