The CB12 product range features two standard versions, which are ideal for a vast number of medical
and industrial applications.
In general the CB12 is a transformer operated control unit, which can control up to 4 actuators.
The control box feature a range of built-in safety devices, increased current cut-off, EAS (Electronic
Arc Suppression), and other options such as battery backup, earth outlet, wet alarm etc.
CB12 with battery backup
The CB12 with battery backup has an acoustic alarm, which sounds when the batteries are low,
approx. 17 - 18V. To charge the batteries on a CB12 with internal batteries, just connect the CB12 to
mains. With external batteries, connect the external batteries to an external charger.
External battery charger
If anything other than a LINAK
®
charger is used, it must conform to the following specifications:
Charging voltage: 27.6 V DC ± 2%
Charging current: < 300 mA.
Warning
In order to avoid injury, a system with control(s) or accessories, a CB12 with battery backup and actuators assembled, must always be
disassembled in transport and service situations.
10. CB12 (MEDLINE
®
CARELINE
®
)
CB20 is a platform which offers a unique safety concept, logging off service data and it is possible to
connect a variety of accessories to the control box.
The CB20 consists of 3 modules :
CP20 = Control Power CU20 = Control Unit BA20 = Battery.
Usage:
• Compatible with specific versions of LA23, LA31, LA34, LA40, LA43, LA44 and BL1
Duty cycle: 10 %
˜
max. 2 min. continuous use followed by 18 min. not in use
Ambient temperature: + 5 ˚C to + 40 ˚C
• Approvals: IEC60601-1:2005 3
rd
ed., ANSI / AAMI ES60601-1:2005, 3
rd
edition, CAN/CSA-22.2 No 60601-1:2008 approved
Please be aware:
CB20 is delivered in 3 separate units - CU20 + CP20 + BA20. The units are not assembled at LINAK A/S.
Microprocessor
All control boxes with a microprocessor must be initialized before start-up. A description of the initialization procedure can be obtained from your
LINAK dealer. If an actuator is replaced, the microprocessor always has to be initialized before use (actuators with reed/hall). If re-programmed,
please ensure that the correct software is used.
External battery charger
If anything other than a LINAK
®
charger is used, it must conform to the following specifications: Charging voltage: 28.0 VDC ± 2 %
Charging current: < 300 mA.
For recommendations on maintenance and storage of the BA20 battery packs,
please refer to chapter 5.