User's Manual

Page 32 of 44
Connection of emergency lowering:
Please observe that if you order the TD3 with 2 x 9V batteries mounted, the batteries will not be
connected. The customer needs to connect the cable from the battery drawer to the TD3 when installing
the bed and connecting the mains.
If the batteries are connected and the bed is not connected to the mains, the batteries will discharge in
approx. 2 months and it will not be possible to perform an emergency lowering.
Lifetime of batteries:
The batteries must be changed after an emergency lowering has been performed. Batteries that have
not been used must be changed after approx. 3 years.
To pull out the batteries
press here
Remember to connect the
cable again afterwards
Battery type:
Alkaline 9 V batteries 9F22 or 6LR61.
The correct way to use the functionality:
If the key is released during the lowering, always wait for 60 seconds before pressing the key again.
Removal/fitting of the 2 x 9 V batteries:
To change the 2 x 9 V batteries for emergency lowering the first step is to disconnect the cable and then pull
out the drawer of the TD3. On the bottom of the drawer, there are 2 holes, which are to be pushed with a
finger or a blunt instrument. The batteries will be released from the drawer when pushing. The new batteries
can be placed in the drawer. Afterwards, place the drawer in the TD3 and connect the cable again.
Only use alkaline batteries, and make sure to polarise the batteries correct according to the symbols in the
drawer.