Datasheet

LTC2414/LTC2418
6
241418fa
Note 1: Absolute Maximum Ratings are those values beyond which the
life of the device may be impaired.
Note 2: All voltage values are with respect to GND.
Note 3: V
CC
= 2.7V to 5.5V unless otherwise specified.
V
REF
= REF
+
– REF
, V
REFCM
= (REF
+
+ REF
)/2; V
IN
= IN
+
– IN
,
V
INCM
= (IN
+
+ IN
)/2, IN
+
and IN
are defined as the selected positive
and negative input respectively.
Note 4: F
O
pin tied to GND or to V
CC
or to external conversion clock
source with f
EOSC
= 153600Hz unless otherwise specified.
Note 5: Guaranteed by design, not subject to test.
Note 6: Integral nonlinearity is defined as the deviation of a code from
a straight line passing through the actual endpoints of the transfer
curve. The deviation is measured from the center of the quantization
band.
Note 7: F
O
= 0V (internal oscillator) or f
EOSC
= 153600Hz ±2%
(external oscillator).
Note 8: F
O
= V
CC
(internal oscillator) or f
EOSC
= 128000Hz ±2%
(external oscillator).
Note 9: The converter is in external SCK mode of operation such that
the SCK pin is used as digital input. The frequency of the clock signal
driving SCK during the data output is f
ESCK
and is expressed in kHz.
Note 10: The converter is in internal SCK mode of operation such that
the SCK pin is used as digital output. In this mode of operation the
SCK pin has a total equivalent load capacitance C
LOAD
= 20pF.
Note 11: The external oscillator is connected to the F
O
pin. The external
oscillator frequency, f
EOSC
, is expressed in kHz.
Note 12: The converter uses the internal oscillator.
F
O
= 0V or F
O
= V
CC
.
Note 13: The output noise includes the contribution of the internal
calibration operations.
Note 14: Guaranteed by design and test correlation.
SYMBOL PARAMETER CONDITIONS MIN TYP MAX UNITS
t
1
CS to SDO Low 0 200 ns
t2 CS to SDO High Z 0 200 ns
t3 CS to SCK (Note 10) 0 200 ns
t4 CS to SCK (Note 9) 50 ns
t
KQMAX
SCK to SDO Valid 220 ns
t
KQMIN
SDO Hold After SCK (Note 5) 15 ns
t
5
SCK Set-Up Before CS 50 ns
t
6
SCK Hold After CS 50 ns
t
7
SDI Setup Before SCK (Note 5) 100 ns
t
8
SDI Hold After SCK (Note 5) 100 ns
The denotes specifications which apply over the full operating temperature
range, otherwise specifications are at T
A
= 25°C. (Note 3)
TI I G CHARACTERISTICS