Setup & Op User guide

30

VALUES ARE FOR PROVING USE ONLY. NOT TO BE USED AS ACTUAL K-FACTORS.
LC Meter Size  Revs / Gallon Pulses / Gallon Pulses / Litre
P
60 12.237 4894.8 1293.11
M/MA 5
60 4.079 1631.6 431.04
M/MA 7/10
150 5.555 2222.0 587.01
M/MA 15/25
300 2.058 823.2 217.47
M 30/40
450 0.742 296.8 78.41
M 60/80
800 0.398 159.2 42.06
MS-75
700 0.255 102.0 26.95
MS-120
1200 0.158 63.2 16.70
CALIBRATION SETUP
To prove a meter system to a product with multi-

1. With the selector switch in the calibration position,
navigate to the Calibration Setup 1 screen.
2. Use the instructions on the page 29 to run a single point
calibration.
3. Navigate to Calibration Setup 3. Move the pointer to
Linearization Mode: and select Setup.
4. Move the pointer to the BEGIN CALIBRATION
DELIVERY eld.
5. Press the EntEr button to begin the calibration delivery
into a volumetric prover. A delivery screen will appear that
shows the gross volume and ow rate.
6. To end the delivery, move the pointer to the END
CALIBRATION DELIVERY eld at the bottom of the
calibration delivery screen.
7. Move the pointer to the Prover Quantity: and
enter the quantity indicated by the volumetric prover.
8. The LCR 600 will automatically display a Point #: list
box with the numbers 1-10. Select a number.
9. The LCR 600 will then display a Linearization Flow
Rate eld edit box. Enter the ow rate of the proving run.
The percent error is calculated automatically.
Multi-Point Calibration Proving
Multi-point calibration improves the overall meter accuracy by attening out the accuracy curve across ow rates.
Although accuracy tends to lessen at lower ow rates, the repeatability of the meter remains consistent. The meter
might not be perfectly accurate at low ow rates, but it is inaccurate the same amount each time. Multi-point calibration
takes advantage of the steadfast repeatability of a Liquid Controls meter by identifying the amount of inaccuracy and
correcting it with a linearizing algorithm during deliveries. In order to apply the linearizing algorithm, the degree of error
for points along the accuracy curve must be identied by making a number of deliveries at different ow rates into a
volumetric prover. Multi-point calibration is very benecial for meter systems that experience a wide range of ow rates
(for example, trucks that ll tanks of varying sizes) and for meter systems that have recently undergone maintenance
or other alterations that could change the accuracy curve itself.
Run the rst multi-point calibration delivery at the same
ow rate used in step 2.
Keep Flow Rates Even
For the rst two proving runs, one should be near the maximum
ow rate and one near the minimum ow rate of the meter.
When choosing ow rates after the initial two proving runs, split
the difference until all points are within 0.25% of each other.
Multi-Point Calibration Tip
Proving runs are in gross volume only.