CALIBRATION SETUP
Multi-Point Calibration Proving - continued
LCR-600: 60011
Adjacent linearization points out of
0.25% range.
The LCR 600 will not allow you to enter the Applied
setting of the Linearization Mode: until all adjacent
calibration points are within 0.25% of each other.
To prove a meter system to a product with multi-

10. Repeat steps 4 through 10 using a new ow rate for the
next Point #: until all the points between the lowest
ow rate and the maximum ow rate are within 0.25% of
each other.
11. Navigate to Calibration Setup 1. Move the pointer to
Linearization Mode: and select Applied.
The linearization points on Calibration
Setup 3 will automatically sort by ow rate
after the Applied setting is selected.